2.3.6
NGA CASTORE XL iQ
20.40
NGA CASTORE XL iQ 20.40
Models: 6920.70.21
-
.1
N2 outlet
Flow rate (Max)
20 l/min
Outlet pressure (Max)
8 Bar (116 psi)
Nitrogen purity
LCMS grade (up to 99.9%
*1
)
Nitrogen for collision cell
*2
200 cc/min @ 99.9999%
*3
with option
6920.71.90
Dry Air outlet
Flow rate (Max)
40 l/min
Outlet pressure (Max)
8 Bar (116 psi)
Dew point
*4
< -40°C (- 40°F)
Communication
LCD with touch screen
Standard
RS485
Standard
RS232
For service
General data
Power supply voltage (min-Max)
220-240 Vac ±10% 50/60 Hz
115-240 Vac ±10% 50/60 Hz
Connection type
IEC C20
Nominal power (Max)
1.8 KW
1.5 KW
Net weight
180 kg
190 kg
Noise level
< 50 dB
Unit fuse F1 (10x38mm)
15A (600Vac – T – class CC)
15A (600Vac – T – class CC)
Unit fuse F2 (10x38mm)
6A (600Vac – T – class CC)
-
Inverter protection fuses F3 - F4
(10x38mm)
-
15A (300Vdc – T – class CC)
Main-board fuse F5 (5x20mm)
3.15A (250Vac – T – L / 35 A)
Heat value (BTU)
6150
5130
Dimensions (W x D x H)
59 x 92 x 73 cm
Connections
N2 outlet port
¼" BSPP female
Dry Air outlet port
¼" BSPP female
N2 UHP outlet port
¼" BSPP female
Drain port
¼" BSPP female
*1
The purity refers to the residual oxygen, depends on flow and working pressure
*2
Optional
*3
The purity refers to the residual oxygen
*4
Atmospheric Dew Point (ADP)
Operating manual NG(A) CASTORE XL iQ
10