OPERATING MANUAL 

                                                   SONIMIX3030-PERM

1T1PT

Edition 1.0E   January 2003

 Page 3

OMSX3030-PERM1T1PT.1oc   10.01.2003

Main technical data

Air       

: Dry air coming

From a gas cylinder
From a Zero Air Generator LNI_SONIMIX

Concentration  @ 5 Nl/min 

: to be choose between  10 ppb and 500 ppb

Concentration range

: 10x by flow variation (0.5Nl/min à 5 Nl/min)

(only with SX 3030-perm1T-Mpt )

Maximal flow

: 5000 Nml/min of mixture, (equivalent N2)

(other flow available)

Precision of the concentration     

: SX 3030-Perm-1pt

 better than 0.6% relative

(for SO2, NO2 and BTEX only)

  

SX 3030-Perm-Mpts 

better than 1 % relative

Repeatability                      

: SX  3030-Perm-1pt

 better than 0.3% relative

  

  SX 3030-Perm-Mpts 

better than 0.5% relative

Stabilisation time of the mixture

:

 < 1 minute

Inlet pressure

         

: from 2.2 to 3.5 bar relative

Max outlet pressure                

: atmosphere

Alarms                            

: inlet pressure, oven temperature, MFC voltage

Command

     

: electrical, local or remote

* local with key board      

: screen VFD 2x20 characters, key board 18 touches

  * remote command

: RS-232, 1200 à 9600 baud, other parameters 100% adjustable

communication Protocol 

: see options

Electrical consumption           

: 25W 220VAC/50Hz

Leak test 

: 316L < 10-7 mbar l/s, checked by  He detector

Casing                            

: ½ 19", 3U, 500mm

SX 3030-Perm-Mpt

No d’article

Option

Descriptif

670210300

SONIMIX 3030-Perm-Mpt

Module à perméation multi-points avec 1 MFC / boîtier ½ 19’

670210302

SONIMIX 3030-Perm-Mpt-OP1

Outlet 3/2 valve to select Cal / Sample

670210311

SONIMIX 3030- Perm -Mpt-OP9.1

Remote command RS 232 : Protocol ASCII

670210312

SONIMIX 3030- Perm -Mpt-OP9.2

Remote command  RS 232: Protocol Breitfuss

670210313

SONIMIX 3030- Perm -Mpt-OP9.3

Remote command  RS 232: Protocol Bayer-Hess

670210397

SONIMIX 3030- Perm -Mpt OP97

Voltage 230 VAC / 50 Hz

670210399

SONIMIX 3030- Perm -Mpt OP99

Voltage 117 VAC / 60 Hz

SX 3030-Perm-1pt

 

No d’article

Option

Descriptif

670210600

SONIMIX 3030- Perm -1pt

Module à perméation mono-point avec BS / boîtier ½ 19’

670210602

SONIMIX 3030- Perm -1pt -OP1

Outlet 3/2 valve to select Cal / Sample

670210611

SONIMIX 3030- Perm -1pt -OP9.1

Remote command RS 232 : Protocol ASCII

670210612

SONIMIX 3030- Perm -1pt -OP9.2

Remote command  RS 232: Protocol Breitfuss

670210613

SONIMIX 3030- Perm -1pt -OP9.3

Remote command  RS 232: Protocol Bayer-Hess

670210697

SONIMIX 3030- Perm -1pt OP97

Voltage 230 VAC / 50 Hz

670210699

SONIMIX 3030- Perm -1pt OP99

Voltage 117 VAC / 60 Hz

No d’article

Accessoires :

670210X72

Tube à permeation SO2

670210X74

Tube à permeation NO2

670210X76

Tube à permeation H2S

670210X78

Tube à permeation NH3

Содержание 670210300

Страница 1: ...HE INSTRUMENT 2 3 INTERNAL STRUCTURES 2 4 ELECTRONICS 3 SOFTWARE DESCRIPTION AND ACCESSIBLE FUNCTIONS 3 1 INTRODUCTION 3 2 SOFTWARE EXPLORATION AND DEFINITION OF THE PRINCIPAL KEYS 3 3 DETAILS OF THE...

Страница 2: ...ts MFC sonic Nozzles pressure regulators permeation oven and ozone generator have been developed by LN INDUSTRIES SA Following features have been in the centre of the development High precision and st...

Страница 3: ...mbar l s checked by He detector Casing 19 3U 500mm SX 3030 Perm Mpt No d article Option Descriptif 670210300 SONIMIX 3030 Perm Mpt Module perm ation multi points avec 1 MFC bo tier 19 670210302 SONIM...

Страница 4: ...bration In Local mode the interface allows to Configure the calibrator and to perform the manual operations of calibrations necessary after the maintenance process on the monitor The major parameters...

Страница 5: ...ubes 30 g m 3 Benzen 30 g m 3 Toluen 30 g m 3 O xylen 30 g m 3 M xylen 30 g m 3 P xylen 30 g m 3 Ethyl Benzen Total outflow of mixture 1000 Nml min Permeation tubes protection flow 100 Nml min Stabili...

Страница 6: ...rding to CE recommandations Power consumption 110V 60Hz or 230V 50Hz 60Hz 150W Gas connections by 1 4 Swagelok fittings Dimentions and weight 19 3HE 84TE 500mm deep About 18 Kg 2 2 PRESENTATION OF THE...

Страница 7: ...rotected by two 1A fuses with a sticker indicating the voltage 2 2 3 Rear facewith Option1 The rear face is composed of 1 the Outlet of mixture fitting Swagelok PVDF1 4 to be connected to analyzers 2...

Страница 8: ...EV6 EV7 EV8 EV9 FILTER EV10 RES1 RES2 FILTER FILTER PS1 PS2 RP1 RP2 DRYER 2 3 2 FLUIDIC MODULES The fluidic circuit of the SONIMIX 6000D1 shows four modules the mass flow generator based on a high pre...

Страница 9: ...tration elastomer seals are made of high grade Viton b Permeation oven module all metallic part in contact with the high concentration mixtures are made of non reactive corrosion free material such as...

Страница 10: ...ards 1 2 and 3 are fixed together on an L shape aluminia cooler located close to the display The boards 4 and 5 are fixed on the front face The three main boards MB3000 PS3000 and FOUR P6L2 and eventu...

Страница 11: ...After cold booting or a RESET RST the instrument identify it self during 5 seconds and shows S O N I M I X 6 0 0 0 D 1 B T X C A L I B R A T O R and jumps to the SAMPLE status S A M P L E In order to...

Страница 12: ...w CST2 100 Ncc The informations are accessible by selecting first the Features menu and after with the and key to read the data 3 3 2 SETTINGS The SETTINGS menu is intended to consult or modify the de...

Страница 13: ...N G S P E R M E A T I O N T U B E 1 Press ENTER to select the right tube 2 for example S E L E C T I O N G A S B E N Z E N E Press on or to chose the gas contained in the tube 2 and then Enter to sele...

Страница 14: ...ample PT3 Press on ENTER P T 2 T O L U E N E A M O U N T 1 9 p p b The stored value in our example 19 ppb is issued of a measurement made with a monitor and compared with a reference mixture at a simi...

Страница 15: ...election by ENTER M A N U A L C O N T R O L S A M P L E M A N U A L C O N T R O L Z E R O M A N U A L C O N T R O L S P A N When SAMPLE is selected S A M P L E 0 0 2 3 The display shows the function e...

Страница 16: ...O 0 0 5 7 S P A N 0 2 3 6 IMPORTANT NOTICE At any time during a zero or span phase the functioning parameters assembled in the BOARD PANEL menu could be monitored by pressing ENTER To exit from the B...

Страница 17: ...H E A T T E M P E R A T U R E T M I N 4 4 5 C T P H E A T T E M P E R A T U R E T M A X 4 5 5 C Press MENU to reach the next active parameter 3 3 5 3 P Tube calibration values B O A R D P A N E L S T...

Страница 18: ...ruption in the wiring or a damaged board 3 3 6 2 AIR Alarm The alimentation of Air pressure from the Zero Air Generator or from a gas cylinder is not setted at a correct value It is inferior to the mi...

Страница 19: ...AIR supply with a stainless steel tube specially processed for chromatography use In addition the pressure reducers fixed on the cylinders must be of good quality and cleaned made of stainless steel...

Страница 20: ...ures superior to 5 bar even temporarly may damage definitively the internal pressure regulators 4 1 2 REMOTE COMMAND CONNECTIONS The remote control is made through a digital interface with only 3 sign...

Страница 21: ...OPERATING MANUAL SONIMIX3030 PERM 1T1PT Edition 1 0E January 2003 Page 21 OMSX3030 PERM1T1PT 1oc 10 01 2003...

Страница 22: ...ERMEATION OVENS TUBES The permeation oven is equiped with high quality long life permeation tubes their life time is expected of about 12 months After 8 to 10 months certain permeation tubes could sho...

Страница 23: ...e between 2 5 and 3bar relative Verify the stabilized AIR pressure at the pressure regulator RP1 s outlet If one of the pressures is inferior to the limit values verify the alimenting pressures and in...

Страница 24: ...24 OMSX3030 PERM1T1PT 1oc 10 01 2003 5 3 5 1 Processor board MB 3000 5 3 5 2 Power supply PS3000A TEST POINT or NAME VALUE Vcc 5VDC U2 Uo U1 Vcc GND U2 Uo U1 Vcc GND PROCESSOR INPUT ALARMS PS1 AND PS2...

Страница 25: ...10 01 2003 GND 0 V V 16VDC without load 12 VDC on load V 8VDC without load 6 VDC on load Vb COMMON 5 3 5 3 KEYBOARD The 18 keys soft keyboard 5 3 5 4 VFD Display The 2X20 characters Vacuum Fluorescen...

Страница 26: ...cessary to maintain the integrity of the product and the validity of the warranty 5 5 FACTORY RECALIBRATION The SONIMIX 6000D1 is a reference intrument which was carefully calibrated with high precisi...

Страница 27: ...quashed 1 Verify the consumption of your monitor during calibration and zero phases 2 Verify the tightness of the circuit 3 Verify all the fittings and tubings between the calibrator and the monitors...

Страница 28: ...iode connection sense No message on the display when starting the instrument 1 One fuse or both are blowed 1 Change the fuse s If the fuse fails repeatedly the problem is probably more important Verif...

Страница 29: ...OPERATING MANUAL SONIMIX3030 PERM 1T1PT Edition 1 0E January 2003 Page 29 OMSX3030 PERM1T1PT 1oc 10 01 2003 NOTES Following...

Отзывы: