Warning! 

Do not remove the vessel from the heating plate when the instrument is working. 

 
Heating function 

The  device  is  controlled  by  digital  temperature  control  technology  which  has  two  separate  safety  

circuits. The heating plate is kept at a constant temperature by a digital control circuit. For a precise 

display and control of temperature inside the medium, please use the external temperature sensor  

PT 1000.  
 
• 

Plug in the external PT 1000. 

• 

Set the temperature by turning the temperature control knob slowly to the target value. 

• 

When the heating function is switched on, the LED displays the temperature value on the left- 

hand side. 

• 

The heating function is switched on or off by pushing the heating knob. 

 
When turned on, the instrument automatically displays the last running speed and temperature para- 

meters. Generally, the LED screen cannot display the actual temperature of the sample in the vessel 

or of the surface of the heating plate. The temperature differences are as following: 
 
• 

Heating plate center and outer edge. 

• 

The sample in the vessel and the surface of the heating plate. 

In  order  to  ensure  the  accuracy  of  the  temperature  inside  the  container,  please  use  the  external 

temperature  sensor  PT1000. 
 
 

Using an external temperature sensor PT 1000 

When the PT 1000 sensor is connected and the temperature control knob is rotated, the LED displays  

the temperature setting value and shifts to real value in 5 seconds. Compared  with the temperature 

control of the heating plate, the external temperature sensor can control the medium’s temperature  

more precisely. The heating function will be stopped automatically under abnormal conditions. 

 
Please operate following the instructions below: 
 
• 

Switch  OFF  the  instrument. 

• 

Ensure the external temperature sensor is inserted into the heated media. 

• 

Switch ON the instrument and run heating function. 

 
If  the  heating  function  does  not  work,  please  contact your distributor. 

 

 

 
 
 

                     

 

 

 

 

      

Содержание uniSTIRRER 3

Страница 1: ...ENG LISH Operation Manual Magnetic stirrer with heating uniSTIRRER 3...

Страница 2: ...result from mechanical vibrations During operation Gradually increase the speed reduce the speed if the stirring bar breaks away due to high speed or the instrument is not running smoothly or the con...

Страница 3: ...t carefully and check for any damages If the instrument is faulty you should contact your distributor The packing includes 1 Magnetic Stirrer with Heating 1 Temperature sensor PT 1000 with stand 1 pow...

Страница 4: ...he speed setting value and flashes The setting value does not flash until real speed reaches the set point Make sure the required operating voltage and power supply voltage match Ensure the socket is...

Страница 5: ...ctual temperature of the sample in the vessel or of the surface of the heating plate The temperature differences are as following Heating plate center and outer edge The sample in the vessel and the s...

Страница 6: ...f 10 rpm Troubleshooting Instrument can not be powered ON check that the power cord is plugged in Fault in power on self test switch OFF the unit then switch ON and reset the instrument to factory def...

Страница 7: ...es if the instrument is opened by untrained or unqualified staff Our General terms and conditions apply Service If help is needed please contact your distributor Please provide the customer care repre...

Страница 8: ...r input 5 W Motor power output 3 W Speed range 100 1500 rpm Display LED Heating plate diameter 135 mm Heating power 500 W Temperature range Room temp to 280 C in 1 C increments Safety temperature 320...

Страница 9: ...Lab Logistics Group GmbH LLG GmbH Am Hambuch 1 D 53340 Meckenheim Tel 49 2225 9211 0 Fax 49 2225 9211 11 www llg de info llg de...