LLG uniSTIRRER 3 Скачать руководство пользователя страница 2

 

1.  Safety instructions 

Before operation 

• 

Read the Instruction Manual carefully before use. 

• 

Ensure that only trained staff use the instrument. The instrument may only be opened by a 

qualified and trained technician. 

• 

Make sure the socket is earthed before use. 

• 

Set the instrument on a stable, clean, dry, non-slip and fireproof surface. 

• 

Do not operate in hazardous areas. 

• 

Before each use, check the instrument for damage. Do  not  use  damaged components. 

• 

The operation voltage noted on the label must match. 

• 

Always disconnect the plug before fitting accessories. 

• 

When working, wear personal protection to avoid the risk from splashing of liquids and 

possible glass breakage which can result from mechanical vibrations. 

 

During operation 

• 

Gradually increase the speed, reduce the speed if: 

-  the stirring bar breaks away due to high speed, or 

-  the instrument is not running smoothly, or  

-  the container moves on the base plate. 

• 

Temperature must always be set at least 25 °C lower than the fire point of the media used. 

• 

Be aware of hazards due to: 

-  flammable materials or media with a low boiling temperature, or 

-  overfilling of media, or  

-  the container is unsafe. 

• 

Do not work with defective parts. 

• 

When the external temperature sensor is required, the tip of the measuring sensor must be at  

least 5 - 10 mm from vessel bottom and wall. 

• 

The instrument can only be disconnected from the main power supply by pulling out the 

main or the connector plug. 

• 

Keep a good, clear distance around the instrument (at least 10 cm from wall/other equipment). 

• 

Keep away from high magnetic field. 

 

 

Содержание uniSTIRRER 3

Страница 1: ...ENG LISH Operation Manual Magnetic stirrer with heating uniSTIRRER 3...

Страница 2: ...result from mechanical vibrations During operation Gradually increase the speed reduce the speed if the stirring bar breaks away due to high speed or the instrument is not running smoothly or the con...

Страница 3: ...t carefully and check for any damages If the instrument is faulty you should contact your distributor The packing includes 1 Magnetic Stirrer with Heating 1 Temperature sensor PT 1000 with stand 1 pow...

Страница 4: ...he speed setting value and flashes The setting value does not flash until real speed reaches the set point Make sure the required operating voltage and power supply voltage match Ensure the socket is...

Страница 5: ...ctual temperature of the sample in the vessel or of the surface of the heating plate The temperature differences are as following Heating plate center and outer edge The sample in the vessel and the s...

Страница 6: ...f 10 rpm Troubleshooting Instrument can not be powered ON check that the power cord is plugged in Fault in power on self test switch OFF the unit then switch ON and reset the instrument to factory def...

Страница 7: ...es if the instrument is opened by untrained or unqualified staff Our General terms and conditions apply Service If help is needed please contact your distributor Please provide the customer care repre...

Страница 8: ...r input 5 W Motor power output 3 W Speed range 100 1500 rpm Display LED Heating plate diameter 135 mm Heating power 500 W Temperature range Room temp to 280 C in 1 C increments Safety temperature 320...

Страница 9: ...Lab Logistics Group GmbH LLG GmbH Am Hambuch 1 D 53340 Meckenheim Tel 49 2225 9211 0 Fax 49 2225 9211 11 www llg de info llg de...