energizing. The picture below shows the display interface of the negative pressure
gauge.
3.
Press “MODE” button to enter OUT2 setting, the main display shows OFF for the
first energizing. Through pressing
▲
and
▼
button, to set OUT2 output mode. The
picture below demonstrates display interface of the negative pressure gauge.
4.
Press “MODE” button to enter NO/NC switching interface, and select NO/NC
through
▲
and
▼
buttons. We set output 1 as NO and output 2 as NC. The picture
below shows the display interface of the negative pressure gauge.
5.
Press “MODE” button to enter response time setting, and set default time as 2.5.
The picture below shows the display interface of the negative pressure gauge.
6.
Press “MODE” button to enter color conversion setting of the main display, and
choose the color of main display through
▲
and
▼
buttons.
The picture below
shows the display interface of the negative pressure gauge.
7.
Press “MODE” button to enter unit setting. The default unit of negative pressure
gauge is shown as KPa, which is the unit we need. Unit could be selected through
▲
and
▼
buttons. The interface is shown as the picture.
8. By this time Menu setups have been circled a circle. Press “MODE” button
once again to return. The main display shows actual air pressure values.Or
long press “MODE” button for 2 seconds during the setting to return to the
home page.
(
EASY Mode) (Hysteresis Mode) (Window Comparator Mode)