31
USER GUIDE
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity Test
IEC 60601-1-2
Test level
Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000-4-4
Not applicable
Not applicable
Surge IEC61000-4-5
Not applicable
Not applicable
Voltage dips, short interruptions and voltage
variations on power supply input lines
IEC 61000-4-11
Not applicable
Not applicable
Power frequency magnetic field IEC 61000-4-8
30 A/m 50Hz/60Hz
30 A/m
50Hz/60Hz
Conduced RF IEC61000-4-6
Not applicable
Not applicable
Radiated RF IEC61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE U
T
is the a.c. mains voltage prior to application of the test level.