Troubleshooting
Problem
Possible Reason
Resolution
Device doesn’t power on
when START/STOP button
is pressed.
or
Cuff does not inflate.
(display remains off)
Batteries completely drained.
Replace all batteries with new ones.
Batteries inserted incorrectly.
Remove batteries and reinstall them,
making sure the + and - polarity symbols
on the batteries match those inside the
battery compartment.
Cuff does not inflate.
(display turns on)
Cuff not plugged in correctly.
Make sure the connector on cuff tube is
plugged all the way into the device.
Cuff doesn’t form a loop
around arm.
Cuff not assembled.
To form a loop: Pass the end of the cuff
farthest away from the tubing through
the long metal D-ring. The smooth side
(without the hook and loop fasteners)
should be on the inside of the cuff loop.