If the Unit is ON and is unused for 3 minutes it will
be switched off automatically.
AUTOMATIC RE-INFLATION
When during the fi rst blood pressure measurement
the cuff inflation to a level of 190 mmHg is not
suffi
cient or you move your arm the Device stops
measurement and re-infl ates the cuff to the higher
level. The Device has 4 fi xed levels of the arm cuff
infl ation: 190, 230, 270 and 300 mmHg.
The automatic re-inflation is repeated until the measurement is completed
successfully. This is not a defect.
FORCED DEFLATION FROM A CUFF
For rapid air release from of the arm cuff during arm cuff infl ation or during a
measurement (slow defl ation) press the «START» button. The device will quickly
release all air from cuff and will switch over to mode of date and time indication.
MEMORY FUNCTION
1. Result of each measurement (pressure, pulse, time and date
of measurement) may be kept in the device memory. For
this purpose, after measurement, within not more than 3
minutes, memory M1 or M2 shall be selected for memory
storage МEM1 or МEM2.
IF THE NOTICE ON ERROR APPEARS THE MEASUREMENT
RESULT WILL NOT BE STORED
.
2. Up to 60 measurement results and mean value of last 3
measurements may be kept in each device memory. When
the number of measurements exceeds 60, the oldest record is
deleted to save the most recent values.
3. Press the МEM1 or МEM2 to see the fi gures stored in the memory. At the fi rst
depression of button МEM1 (or МEM2) mean value of 3 last measurements,
kept in memory M1 (or M2), will appear on the screen, marked by symbol « »
(fi g. 17). At repeated depression of button MEM1 (or МEM2) indicator of selected
memory МEM1 « » (or МEM2 « ») and number of memory cell will appear on
the screen, and in 1 second its content is displayed (fi g. 18).