MEMORY CLEARING
To delete all values stored in the memory, press the M
Button and hold it down for more than 5 seconds. Symbols
“Clr” will appear on the screen and all values will be
deleted from the memory (Fig. 8).
INFORMATION ABOUT ERRORS
Indication
Likely cause
Methods of correction
The arm cuff is applied incorrectly
or the air tube plus is inserted not
tightly enough.
Measurements cannot be made
due to hand movement or talking
during measurements.
At such disorders as deep vascular
sclerosis, weak pulse wave and
also in patients with the prominent
distortions of cardiac rhythm it may
be dif
fi
cult to measure the blood
pressure accurately.
Be sure that the arm cuff is
applied correctly and the plug
is inserted tightly. Repeat
the whole measurement
procedure.
Repeat the measurement
following strictly the recom-
mendations of this Manual.
In such cases consult
a certified doctor about
application of the electronic
Device.
Batteries are discharged.
Replace the batteries for new
ones
Fig .8