14
Hardware and Calibration
AmpConnect 621
™
Manual
Listen Hardware Tab
AmpConnect 621 is automatically recognized by SoundCheck according to serial number. Each device is
added to the Listen Hardware page and automatically named. If simply exchanging one AmpConnect 621
for another, the Device ID remains
AC621-1
and the
Serial #
field updates.
Right-click on the
Device ID
to change the following:
Rename
- Change the
Device ID
field to a
unique name. The example in
Figure 4-11
shows the Device ID of AmpConnect 621 is
set to:
AC621-1
Delete
- Remove the selected item from the
list
Assign Startup Default
- Set the initial
settings that the device will adopt when
connected to this SoundCheck system
Startup Default
The startup default editor, shown in
Figure 4-12
, initially contains settings commonly used for loudspeaker
testing, explained below. These can be modified to suit your test requirements.
Note:
The Self Test sequence will return AmpConnect 621 to factory default settings. Restore your
Startup Default settings by clicking
Apply
in the editor show in
Figure 4-12
or restart
SoundCheck. See
Self Test Sequence on page 16
.
Inputs
Source
: All BNC inputs are routed to the their respective Audio
Interface channels, e.g.: BNC Input 1 routed to Audio Interface
Input 1
Gain
: All channels set to Mic Bias set to Voltage with +20 dB of
gain
Mic Bias
: All channels set to SCM for use with Listen’s SCM
reference microphones
Switching Mic Bias on or off with an AmpConnect 621
Message Step during a sequence run may require a
Wait
Time
of 2 seconds in the step to allow for settling time.
Amplifier
: Always routed to Amplifier input of Audio Interface.
Gain is set to 0dB.
Impedance
: Always routed to Impedance of Audio Interface. Gain
is set to 0dB.
Outputs
Output 1 and 2
: The Audio Interface Outputs are routed to the BNC outs respectively. This allows
for the use of an external amplifier if more channels are required.
Amplifier
: Power Amp Output routed to Binding Post Output A
Full details on settings and use of the Message Step are covered in
AmpConnect 621 Message Step on
page 21
.
Figure 4-11: AmpConnect 621 Setup
Figure 4-12: Startup
Default
Содержание AmpConnect621
Страница 1: ...PN 8072 Rev 042622 AMPCONNECT 621 USER MANUAL COPYRIGHT 2020 2021 LISTEN INC...
Страница 2: ...Rev 042622...
Страница 6: ...iv Limited Warranty AmpConnect 621 Manual page intentionally left blank...
Страница 26: ...AmpConnect 621 Manual Hardware and Calibration 20 Page intentionally left blank...
Страница 32: ...26 USB Control Via SoundCheck AmpConnect 621 Manual Page intentionally left blank...
Страница 46: ...40 Equipment Ratings and Approvals AmpConnect 621 Manual CE Declaration of Conformity...
Страница 47: ...AmpConnect 621 Manual Equipment Ratings and Approvals 41 UK Declaration of Conformity...
Страница 48: ...42 Equipment Ratings and Approvals AmpConnect 621 Manual Page intentionally left blank...
Страница 62: ...AmpConnect 621 Manual Listen Software License Agreement 56...
Страница 63: ...AmpConnect 621 Manual Listen Software License Agreement 57 LISTEN Inc 580 Harrison Ave Suite 3W Boston MA 02118...
Страница 64: ...AmpConnect 621 Manual Listen Software License Agreement 58...
Страница 66: ...60 INDEX AmpConnect 621 Manual page intentionally left blank...