5: Warning Messages
5.1 Fatal Error Messages
If a
FATAL
message is shown, the
500
automatically powers down
and cannot be used until the appropriate remedial action has been
taken. The most likely occurring examples of such a message are:
Instrument Too Hot
Instrument Too Cold
Change/Charge Battery
The
Error Code X
messages are explained in the following table. If
any of these occurs, please contact service personnel for attention.
lion alcolmeter
©
500: Error Code Messages
Code
Meaning/Significance
2
Fuel cell amplifier signal is too high
3
Fuel cell output too low during instrument calibration
4
Gas flow rate applied during breath flow sensor calibration was too low
5
Charge pump did not generate the required voltage
6
Temperature below -6°C
7
Calibration or checking attempted below 10°C
8
Calibration or checking attempted above 30°C
16
Fuel cell offset voltage has drifted too low
17
Blockage in sampling system
18
Used in Factory Mode only
19
Used in Factory Mode only
Error: Code 3
Please Refer to
Service Attention
MEMORY FULL!
Out of Use!