lion alcolmeter
®
500 Mk IV
User Handbook
14
3: Subject Breath Test Procedure
3.1 Step by Step …
STEP 1 -
Preliminary Questions to the Subject
As far as possible, ensure the subject has taken
NOTHING
by mouth
[alcoholic or otherwise] in the last
TWENTY
minutes; and has not
smoked in the last
TWO
minutes.
If necessary - WAIT.
STEP 2 -
Switch On
Press
FSA
until you hear the beep, then release. The instrument
serial number is displayed. This is followed by the date and time, and
then a review of the remaining memory capacity:
STEP 3 -
Function Selection
Now scroll through the menu using the small button [
FSA
], to select
what you want to do: there are three options. The menu option line,
shown in these diagrams in
ITALICS,
flashes to indicate that the user
may select this function from the available menu:
Serial Number:
21345
Date: 03/09/2012
Time: 23:56
Remaining Test
Memory: 1389
Subject Test:
NEW TEST
Large Button: SELECT
Small Button: NEXT
Subject Test:
RECALL LAST TEST
Large Button: SELECT
Small Button: NEXT
Subject Test:
SWITCH OFF
Large Button: SELECT
Small Button: NEXT