6
2.
Install the regulator with a 1 1/8 inch
wrench.
NOTE:
The threaded adaptor of the
nitrogen regulator is designed to fit nitrogen
fittings only. Incompatibility of the regulator
and tank indicates a gas source other than
nitrogen or an improper regulator for use
with a nitrogen tank.
3.
Once the regulator is securely installed,
ensure the regulator knob is in the full off
position by turning the regulator control
knob counterclockwise.
SUDDEN
PRESSURE EXERTED TO THE
REGULATOR MAY CAUSE
INTERNAL DAMAGE.
4.
Slowly turn the tank valve fully open
(counterclockwise). This will allow
nitrogen to pressurize the regulator.
5.
Insert the male Schrader end of the hose
into the female Schrader on the regulator
with an upward thrust.
6.
Operating pressure is established by
gradually turning the regulator control knob
clockwise.
ALWAYS establish the
designated pressure on the operating
pressure gauge with the instrument
running.