background image

 

The Hall

 

®

 

 Series 4

 

®

 

Instruction Manual

Содержание Hall Series 4

Страница 1: ...The Hall Series 4 Instruction Manual ...

Страница 2: ...retrieval system or translated into any human or computer language in any form or by any means electronic mechanical magnetic manual or otherwise or disclosed to third parties in whole or in part without the prior express written consent of Linvatec Linvatec reserves the right to revise this publication and to make changes from time to time in the contents hereof without obligation to notify any p...

Страница 3: ...n 8 2 2 1 Drill Reamer 9 2 2 1 1 Connecting Attachments 10 2 2 1 2 Mode Selection and Operation 11 2 2 2 Oscillating Saw 12 2 2 2 1 Blade Attachment Instructions 12 2 2 2 2 Rotating Head Instructions 14 2 2 2 3 Operating Instructions 14 2 2 3 Reciprocating Saw 15 2 2 3 1 Blade Attachment and Operation Instructions 15 2 2 4 Trauma Drill 17 2 2 4 1 Connecting Attachments 17 2 2 4 2 Mode Selection an...

Страница 4: ...n 22 3 1 3 2 Gravity Air Displacement Steam Sterilization 22 3 2 Troubleshooting 23 3 3 Specifications 27 3 3 1 Drill Reamer Handpiece 27 3 3 2 Trauma Drill Handpiece 27 3 3 3 Oscillating Saw 28 3 3 4 Reciprocating Saw 28 3 3 5 Environmental Requirements 29 4 0 CUSTOMER SERVICE and WARRANTY 4 1 Customer Service 30 4 2 Attachments and Accessories 32 4 3 Linvatec and Hall Surgical Instrument Warrant...

Страница 5: ...structions marked on the product and included in this manual 2 Eye protection is recommended when operating equipment 3 Use only Linvatec and Hall accessories and attachments 4 Prior to each use perform the following Inspect all equipment for proper operation Ensure all attachments accessories and hoses are correctly and completely attached to the handpiece Always inspect hoses for signs of wear o...

Страница 6: ...for service 10 Do not pressurize hoses until all fittings have been connected and checked 11 Never operate instrument above 100 psi 7 kg cm2 dynamic pressure unless an extension hose is added to the standard 10 foot hose Excessive pressure may cause damage to instrument and exert unusual stress on the hose 12 Use the appropriate mode selection for the appropriate function Never drive bone screws w...

Страница 7: ...rage area and piped into the operating room Manifold systems are available to eliminate frequent tank changes CAUTION Do not exceed 100 psi 7 kg cm2 operating pressure unless a hose longer than the standard 10 ft Universal Hose 5052 010 or extension hose is used Add an additional 1 psi for every extra foot of hose The Series 4 handpieces should be operated at 100 psi 7 kg cm2 for maximum operating...

Страница 8: ...in the full off position by turning the regulator control knob counterclockwise SUDDEN PRESSURE EXERTED TO THE REGULATOR MAY CAUSE INTERNAL DAMAGE 4 Slowly turn the tank valve fully open counterclockwise This will allow nitrogen to pressurize the regulator 5 Insert the male Schrader end of the hose into the female Schrader on the regulator with an upward thrust 6 Operating pressure is established ...

Страница 9: ...g e The hose can then be removed from the connector Hold the end of the hose securely when disengaging the male Schrader fitting to prevent possible damage to the diffuser 8 If the Hall Pneumatic Connector is being used a Locate the button marked PRESS b Depress and hold the button until the audible release of residual gas is completed c Release the button and remove the hose d If the hose cannot ...

Страница 10: ...wise and slightly pull on the hose so the internal pins securely engage in the indentations 3 Attached to the Universal Hose 5052 010 is the Hall Hose Handler This clip allows fixation of the hose to the surgical drape reducing the problem of the hose slipping off the patient 4 The Velcro strap allows the hose to be coiled thus reducing excess hose length In addition the VELCRO strap may be used w...

Страница 11: ...n Used for Tapping and Setting Bone Screws Direction Control Knob Used to select the operating direction between FWD forward REV reverse and SAFE Place the control knob in the SAFE position when changing accessories To activate the handpiece place the control knob in either FWD or REV Activation Trigger Used to activate the handpiece when the direction control knob is in either FWD or REV Zimmer H...

Страница 12: ...escribed in step 1 AO Shank accessories must utilize the AO adaptor Affix the AO adaptor into the Zimmer Hudson snap lock chuck as described in step 1 1 To connect attachments a Place the Direction Control Knob in the SAFE position b Firmly grasp the chuck and pull back toward the handpiece When inserting Trinkle or AO products care should be taken to pull back on the adaptor itself not the handpi...

Страница 13: ... control handle when using heavy reamers or attachments The torque control handle will assure maximum control of the handpiece 2 2 1 2 Mode Selection and Operation WARNINGS 1 Always make sure you are in the proper mode before the handpiece is activated 2 Never set bone screws with the handpiece in the REAM position Only use the SCREW position for setting bone screws 1 Twist the mode selector ring ...

Страница 14: ... any of 12 positions Rotating Head Houses the cutting blade and rotates to any of 12 positions at 30 intervals for appropriate surgical access Blade Locking Collet Holds the blade in place Blade Locking Knob Locks the blade securely in place Activation Trigger Press to operate the handpiece when the control knob is in the ON position 2 2 2 1 Blade Attachment Instructions 1 To affix a blade to the ...

Страница 15: ... Insert the blade at the desired angle and align the blade holes with the blade positioning pins Be sure that the proper saw blades are used with the collet Micro Sagittal Saw Blades may not be used in the wrenchless collet only Hall series 5071 xxx blades can be used d Rotate the blade locking knob to the closed position ...

Страница 16: ...positions and cutting planes a Grasp the head locking collar firmly and pull towards the back of the handpiece b Rotate the head and blade to the desired position Release the head locking collar 2 2 2 3 Operating Instructions 1 To operate the handpiece a Twist the control knob from the SAFE position to the ON position b Depress the activation trigger ...

Страница 17: ...o activate the handpiece Place in the SAFE position before changing blades bits accessories or hoses and when the instrument is not in use Activation Trigger Press to operate the handpiece when the control knob is in the ON position 2 2 3 1 Blade Attachment and Operation Instructions 1 To affix a blade to the handpiece a Place the control knob in the SAFE position ...

Страница 18: ...ocked in any position Four 4 detents are provided for accurate 90 positioning Before locking the collet grasp the blade at the base of the collet and turn to the desired position e Turn the collet locking knob counter clockwise to firmly secure the blade 2 To operate the handpiece a Twist the control knob from the SAFE position to the ON position b Depress the activation trigger ...

Страница 19: ... Rotate to the vertical position to activate the handpiece Direction Control Lever Used to select the operating direction between FWD forward SCREW and REV reverse 2 2 4 1 Connecting Attachments 1 Place handpiece in the SAFE position a Rotate the trigger either left or right Attachment of Trinkle Shank Accessories 1 Pull back the outer collet sleeve 2 Insert the Trinkle shank into the collet The s...

Страница 20: ... shank with the black arrow on the collet sleeve and insert the accessory into the collet 2 Insert the shank as far as it will go without pulling back the collet sleeve Twist the shank to make sure it is aligned properly If it is it will not spin in the collet 3 Pull back on the collet sleeve while pushing the shank further into the collet 4 Release the collet sleeve The shank is now locked into t...

Страница 21: ... to the desired position NOTE This can be done using the small finger while holding the instrument 2 To select the SCREW mode place the direction control lever in the center there is a detent This automatically limits the torque in the forward mode to 18 22 in lbs for setting bone screws 3 To operate the handpiece depress the trigger ...

Страница 22: ...b the handpiece and attachments with a soft brush and mild detergent For attachments use a nylon brush or pipe cleaner and clean the insides thoroughly Remove all traces of blood coagulated material stains etc DO NOT IMMERSE handpieces or attachments in soap solution or rinse water 3 Keeping the nose of the handpiece pointed downward and with the hose still attached rinse under running water to re...

Страница 23: ...ed for complete heat and moisture dissipation Operation of a handpiece that is not completely cool or dry may decrease performance and or reliability WARNING Use of disinfecting solutions for an exterior instrument wipe will not sterilize the instrument and cannot be recommended 1 Place the cleaned instruments in an instrument tray 2 Remove the hose from the handpiece Do not crimp the hose when cl...

Страница 24: ...avity Air Displacement Steam Sterilization Temperature 270 272 F 132 133 C Exposure time 35 minutes Temperature 250 254 F 121 123 C Exposure time 60 minutes Drying time 8 minutes minimum Wrapped or unwrapped For Sterilization Containers with filters add FIVE 5 minutes to the Gravity cycle NOTE Dry times are extremely important for reliability and durability of handpieces If eight minutes in your a...

Страница 25: ...eing performed Run handpiece on another regulator to see if the problem is the handpiece or regulator Replace appropriate piece of equipment Set pressure to recommended operating pressure If using a hose longer than 10 ft or an extension hose is being used add an additional one psi of pressure per each additional foot of hose Check all hose connections and ensure they are completely seated Remove ...

Страница 26: ...lector ring will not shift from DRILL to REAM or REAM to DRILL With the direction control knob in the FWD position place the mode selector ring in whichever position it will easily go into Briefly activate the handpiece Shift the mode selector ring to the desired position Motor operates but the output shaft does not Direction control knob in the SAFE position Place the direction control knob in ei...

Страница 27: ... Detent on attachment is not aligned with the arrow on the collet Align detent of attachment with the arrow on the collet Handpiece stalls while driving Direction control lever possibly in the SCREW position Place the direction control lever in the FWD or REV position Lack of handpiece power See Drill Reamer troubleshooting information on the previous pages Table 1 Troubleshooting Guide Symptom Po...

Страница 28: ...t in a positive detent hold Blade is not fully seated Place blade in a positive detent position Completely seat blade in collet Control knob will not go into the SAFE position With a light pressure pull the trigger out to its stop Rotate the control knob to the SAFE position Do not force Lack of handpiece power See Drill Reamer troubleshooting information on the previous page Table 1 Troubleshooti...

Страница 29: ...r Handpiece Operating Speed forward and reverse Drill Position 750 rpm Screw Position 250 rpm Ream Position 250 rpm Average Output Torque forward and reverse Drill Position 40 in lbs Screw Position 18 22 in lbs Automatic torque limiter Ream Position 100 in lbs Weight 39 7 oz 1 13 kg Operating Pressure 100 psi running 7 kg cm2 Recommended Power Source Medical grade water pumped compressed dry nitro...

Страница 30: ...ng Pressure 100 psi running 7 kg cm2 Recommended Power Source Medical grade water pumped compressed dry nitrogen Consumption 17 0 cfm maximum 481 L min 3 3 3 Oscillating Saw Operating Speed 14 000 Cycles minute nominal 28 000 Strokes minute nominal Oscillation 5 Arc Blade Rotation 12 positions 30 intervals Weight 33 8 oz 967 6 g Operating Pressure 100 psi running 7 kg cm2 Recommended Power Source ...

Страница 31: ...ed dry nitrogen Consumption 13 0 cfm maximum 368 L min 3 3 5 Environmental Requirements Operating Ambient Operating Temperature 50 F to 77 F 10 C to 25 C Relative Humidity 30 to 75 Atmospheric Pressure 700hPa to 1060 hPa Storage Ambient Temperature 40 F to 158 F 40 C to 70 C Relative Humidity 10 to 100 Atmospheric Pressure 500hPa to 1060 hPa NOTE There are no toxic components used in the manufactu...

Страница 32: ...erial Number Reason for Return Original Invoice Number Date of Purchase Repairs Products returned for repair must have an authorized Return Goods R G number prominently displayed on the box and included on all paperwork Refer to this number if making inquiries about the repair status Please call Linvatec Customer Service and provide the following information to obtain an R G number prior to return...

Страница 33: ...cent 15 of the purchase price minimum charge 25 Products returned as a result of errors attributable to Linvatec are exempt from this fee Returns must have an authorized Return Goods R G number prominently displayed on the box and included on all paperwork Returns must be shipped prepaid freight otherwise they will not be accepted Products must be decontaminated and sterilized before returning Tho...

Страница 34: ...le Adaptor 5044 006 Series 3 4 AO Adaptor 5044 009 Series 3 4 5 32 inch to Jacobs Adaptor 5044 011 Series 3 4 1 4 inch to Jacobs Adaptor 1365 042 Zimmer to Hudson Adaptor 1384 028 Zimmer to Trinkle Adaptor 1384 029 Zimmer to 1 4 inch Jacobs Adaptor 1384 034 Zimmer to AO Adaptor 1384 073 Zimmer to Stryker Adaptor 5052 020 Two Way Pin Driver Zimmer Oscillating Saw REF Description 5067 002 Oscillator...

Страница 35: ...Includes Universal Hose 5067 094 Trauma Drill without Hose 1368 005 Trinkle to 5 32 inch Jacobs Adaptor 1368 010 Trinkle to 1 4 inch Jacobs Adaptor 5040 011 Wiredriver Attachment 5052 019 Two Way Pin Driver Trinkle 5053 123 Wire Guard 5053 124 Cleaning Brush Miscellaneous Handpiece Accessories REF Description 5052 010 Universal Hose 5067 501 Extended Warranty U S only ...

Страница 36: ... SLG Instruments Lifetime 13 Shutt Non SLG Instruments Twelve 12 Months 14 Footswitches Twelve 12 Months 15 Irrigation Systems Twelve 12 Months 16 Reusable Procedure Specific Instruments Twelve 12 Months Linvatec recommends that the Hall Series 4 Handpieces and pneumatic regulators described in this manual be returned to the factory for routine maintenance every twelve 12 months Failure to follow ...

Страница 37: ...y were designed and manufactured including use in any way inconsistent with the instructions and warnings contained in the Company instruction manuals and package inserts 5 Instruments which were last serviced refurbished reprocessed or reconditioned by a nonauthorized service entity 6 Instruments which did not have their aforementioned routine maintenance schedule followed The foregoing limited w...

Страница 38: ...36 ...

Страница 39: ......

Страница 40: ...ev A 01 2000 Hall Surgical 11311 Concept Boulevard Largo Florida 33773 4908 Phone 727 392 6464 Customer Service 800 925 4255 USA Fax 727 399 5256 International Fax 727 397 4540 www linvatec com 2000 Linvatec Corporation a subsidiary of ConMed Corporation ...