Linseis Chip-DSC 10 Скачать руководство пользователя страница 3

How To Low temperature measurements 

 

1.

 

General information 

This manual is a short description for handling measurements with the Chip DSC in combination with a 
cooling option to investigate subambient measurements. The presented options are the most 
common, that are usually used. However, due to continuously development changes and new options 

are possible. For more Information, read the other available instructions or specific manuals for the 
Chip DSC. 

2. Preparation of the Chip-DSC 10 System 

2.1 Peltier system 

o

 

Easy to use and continuous cooling option with thermo electric generators 

o

 

Δ

T up to 30 K below ambient temperature is possible 

o

 

place the cooling block on the ready to use Chip-DSC 10 

o

 

tighten the screws at the blue main plate 

o

 

connect the Peltier cooler with power supply and turn on the cooling device 

2.2 QuenchCooling system 

o

 

Easy to use and efficient cooling option with liquid nitrogen for lowest temperatures 

o

 

Temperatures down to -180 °C are possible 

o

 

Place the cooling block on the ready to use Chip-DSC 10 

o

 

Tighten the screws at the blue main plate 

o

 

Fill the QC container with liquid nitrogen until you reach the required temperature 

o

 

Closing the measurement cell during cooling and measurement avoid condensation near the 

sensor 

 

 

Содержание Chip-DSC 10

Страница 1: ...1 How To Low temperature measurements with Chip DSC Linseis Messgeräte GmbH Gerlach Stand 26 08 2021 ...

Страница 2: ...Preparation of the Chip DSC 10 System 3 2 1 Peltier system 3 2 2 QuenchCooling system 3 3 Preparation of the Chip DSC 100 System 4 3 1 Peltier system 4 3 2 Intracooler System 4 3 3 manual LN2 system 5 3 4 Automatic LN2 system 6 4 Calibration hints 7 5 Measurement hints 7 ...

Страница 3: ...continuous cooling option with thermo electric generators o ΔT up to 30 K below ambient temperature is possible o place the cooling block on the ready to use Chip DSC 10 o tighten the screws at the blue main plate o connect the Peltier cooler with power supply and turn on the cooling device 2 2 QuenchCooling system o Easy to use and efficient cooling option with liquid nitrogen for lowest temperat...

Страница 4: ...ed and can be turned on by the switch at the side of your Chip DSC 100 3 2 Intracooler System o Closed system and easy to use cooling option for investigations at lower temperatures o up to 99 C at coolfinger is possible o insert the coolfinger carefully in the hose of the adapter in your Chip DSC 100 o Tighten the screw at the bottom of the cooling hose o Turn on the Intracooler directly at the b...

Страница 5: ...ns at lowest temperatures o investigations from 150 C are possible o open the cap of your Quench cooling reservoir o Fill the reservoir with liquid nitrogen until you reach the required temperature o To avoid evaporated liquid nitrogen in your lab it is possible to connect an additional tube at the outlet of your cooling option ...

Страница 6: ...est temperatures o investigations from 150 C are possible o adjust the level of the temperature sensor in your liquid nitrogen reservoir o Connect the tube of your Dewar bottle with the inlet of your Chip DSC 100 o To avoid evaporated liquid nitrogen in your lab it is possible to connect an additional tube at the outlet of your cooling option ...

Страница 7: ... t evaporate when you put them from balance to the precooled measurement cell e g by using closed crucibles 5 Measurement hints o Make sure the measurement cell is closed properly to avoid convection and condensation o Using dry gas also helps to avoid condensation also purge while changing the sample o For liquid samples make sure they don t evaporate when you put them from balance to the precool...

Отзывы: