© 2017 Linkam Scientific Instruments Ltd.
18
v1.1.1
5.
Replace the lid of the humidity chamber and place the caps on the white plastic air ports.
6.
Allow the chamber to reach humidity equilibrium (see table below) and switch the RH95 unit on. You can now compare the humidity reading
of the RH95 with the expected humidity value of the calibration standard used to validate the accuracy of the RH95.
Time to reach equilibrium
RH% of humidity standard
45 minutes
10%, 35%, 65%
60 minutes
80%
Note
: After every calibration the textile pad must be renewed and the chamber washed out with water and well dried. Thoroughly wash your hands
to remove any traces of the calibration salts.