Validating the humidity reading

If you have purchased an optional Rotronic calibration standard with your RH95 unit, you can use these to validate the humidity reading of your
RH95. 

You should have received the following package:

This will contain some textile pads and  five ampoules of calibration salts. The RH% will depend upon which kit was ordered. They are available in RH
values of 10%, 35%, 65% and 80%.

These standards consist of salts which will alter the humidity of a small enclosed space to a specific RH value, by either absorbing or releasing
moisture into the air. The humidity readings can be affected by several factors, so the environment must be controlled when using these validation
standards:

·

Room temperature must be 23°C +/-2°C.

·

The humidity standard 

must

 be at the same temperature as the humidity sensor. A 1 degree difference in temperature can cause an error of

up to 6% RH

·

There must be no draughts in the calibration room

·

Do not expose the calibration chamber to direct sunlight while in use

·

Do not expose the calibration salts to direct sunlight at any time.

Usage directions

Note

: It is recommended that you wear safety goggles while performing this validation procedure.

1.

Remove the lid of the supplied humidity chamber and place one of the textile pads in the recess.

Note: 

do not touch the textile pad with bare hands, as this can cause contamination and inaccurate readings. Wear sterile gloves, or use a clean

pair of pliers or tweezers to load the pad.

Содержание RH95

Страница 1: ...RH95 Humidity Controller Linkam Scientific Instruments USER GUIDE 1 1 1...

Страница 2: ...e l appareil 7 System Overview 8 Connecting your RH95 Controller 8 Back Panel Connections 8 USB Connector 8 RS232 Connector 8 Humidity sensor connector 8 Power socket 8 Transit Screw Holder 8 Air Inle...

Страница 3: ...ion Thank you for purchasing your Linkam equipment please take a moment to read this manual before setting up the equipment Ifyou have purchased control software please install the software first See...

Страница 4: ...products that prove to be defective For warranty service or repair this product must be returned to Linkam or a designated service facility The warranty shall not apply to defects resulting from inte...

Страница 5: ...Technical Support 9 The equipment is not intended to be used outdoors 10 Each product is equipped with a 3 wire grounded earth power plug or a free end 3 wire power cord The plug only fits into a grou...

Страница 6: ...produit est quip d une fiche d alimentation mise la terre ou d un cordon d alimentation avec 3 fils extr mit libre La fiche ne s ins re que dans une prise de courant mise la terre Le cordon d alimenta...

Страница 7: ...y chamber 5 Humidity chamber 6 Humidity chamber tool 7 RH water bottle 8 Humidity data RS232 cable only included when a T95 controller is supplied 9 Screwdriver 10 USB Type Ato micro B cable for firmw...

Страница 8: ...nsor connector The humidity sensor connector is used to connect the RH95 to the humidity sensor cable Power socket For connection to the power supply provided with the RH95 unit Note DONOTconnect the...

Страница 9: ...dity sensor isextremely sensitive to contaminants DONOTtouch the exposed chip or allow anything to come into contact with it These sensorscannot be repaired ifbroken or contaminated and are not covere...

Страница 10: ...a Linkam heating stage with humidity capability to go with your RH95 controller follow these instructions to connect it IMPORTANTNOTICE The humidity sensor isextremely sensitive to contaminants DONOTt...

Страница 11: ...ect the humidity tube from the front ofthe RH95 to the port on the Linkam stage as shown 5 Insert the gas valve insert into the gas valve on the opposing side ofthe stage to the humidity feed 6 You ar...

Страница 12: ...d water is used in the Humidity Generator Step 1 Make sure the controller is powered down and that the transit screws have been removed from the base ofthe unit Store these screws in the provided hole...

Страница 13: ...d tighten the locking screw until it is just tight Do not over tighten the screw Step 6 Replace the magnetic front panel ofthe RH95 Important Note Before shipping the RH95 the water bottle should be r...

Страница 14: ...tap the set time display and a keypad screen will appear Enter the time required in hours minutes and seconds and press enter then push the start button to begin the conditioning NOTE The suggested d...

Страница 15: ...mber type it is designed for and the serial number ofthe sensor Touchscreen Calibration Ifthe touchscreen interface on the RH95 is not responding correctly to touch input it may need to be re calibrat...

Страница 16: ...several factors so the environment must be controlled when using these validation standards Room temperature must be 23 C 2 C The humidity standard must be at the same temperature as the humidity sen...

Страница 17: ...at the marked place and empty it completely on the center ofthe textile pad while tapping it lightly to make sure all of the liquid is soaked into the pad Note wrap the top ofthe ampoule with tissue p...

Страница 18: ...nit on You can now compare the humidity reading ofthe RH95 with the expected humidity value ofthe calibration standard used to validate the accuracy ofthe RH95 Time to reach equilibrium RH ofhumidity...

Страница 19: ...ntroller has been found to comply with the following applicable European Directives and carries the CEmarking accordingly EMC Directive 2014 30 EU using product standard EN 61326 1 2013 Low Voltage Di...

Страница 20: ...voltage 24Vdc Max current 3A Display Resolution 128 X 64 pixels Display Size 2 3 inch User Interface Touch screen RH Sensor Digital capacitive sensor RH Range 10 to 90 RH Resolution 0 1 resolution RH...

Страница 21: ...2017 Linkam Scientific Instruments Ltd 21 v1 1 1...

Страница 22: ...Linkam Scientific Instruments Ltd Tel 44 0 1737 363 476 Fax 44 0 1737 363 480 Email support linkam co uk Unit 8 Epsom Downs Metro Centre Waterfield Tadworth Surrey KT20 5LR UK www linkam co uk...

Отзывы: