6.5 Electromagnetic Compatibility (only Sprint 200 with i-Drive Power)
Stretcher is intended for hospitals except for near active HF surgical equipment and the RF shielded room of a medical
system for magnetic resonance imaging, where the intensity of EM disturbances is high.
Stretcher has defined no essential performance.
6.5.1 Manufacturer instructions - electromagnetic emissions
Emission Test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Complies
NOTE
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer
adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as reloca
-
ting or re-orienting the equipment.
WARNING!
It is recommended to avoid the use of this device next to or in block with other device, because it could lead
to improper operation. If such use is needed, this device and the other equipment should be under surveillan-
ce to verify proper operation.
WARNING!
Use of the accessories, converters and other cables, than specified and provided by manufacturer of this
stretcher could lead to increase of electromagnetic emission or lower the electromagnetic immunity of this
stretcher and lead to improper operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna)
should not be used closer than 30 cm (12 inches) from any part of this stetcher Sprint 200 with i-Drive Power,
including cables specified by manufacturer. Otherwise this could lead to deterioration of functionality of this
stretcher.
WARNING!
Do not overload the stretcher (SWL) and consider chapter 20 Maintenance in order to maintain the basic safe-
ty with regard to electromagnetic disturbances for the expected service life.
List of used cables:
►
Mains cable
, maximum length 5 m