35
2) Adjustment by program:
Set-up the instrument on tripod or adjustment platform, and leveling
accurately.
Procedures
1. Power on and enter into INST
Setting. On the screen tap ◄ or
► keys until Collimation display,
tap it to display collimation error
and vertical index error setting
menu.
2. Aim at the cross-hair of collimator at telescope left, tap “Left value” to
read the horizontal and vertical angles.
3. Aim at the cross-hair of collimator at telescope right, tap “Right value” to
read the horizontal and vertical angles.
4. The software will calculate the new collimation error and vertical index
error automatically.
5. Tap “Enter” to save the new values, or tap “Cancel” to use old values.