33
2. Aim at the cross-hair of collimator at telescope left, tap “Left value” to
read the horizontal and vertical angles.
3. Aim at the cross-hair of coillmator at telescope right, tap “Right value” to
read the horzontal and vertical angles.
4. The software will calculate the new collimation error and vertical index
error automatically.
5. Tap “Enter” to save the new values, or tap “Cancel” to use old values.
Note:
The adjustment can be performed by the program when C<30″, if C>30″,
adjust the reticle.
Reticle Adjusting:
1. Rotate the instrument in face right position, turning horizontal tangent
screw until Hr′=Hr+C.
2.
Loosen the shield of telescope’s reticle.
3. Adjusting two screws at left and at right until the vertical hairs of
telescope’s reticle coincides with the cross-hairs of collimator or target.
4. Repeat the check and adjustment procedure until the error is accepted.
NOTE:
1. When adjust the screws of reticle, firstly loosen the screw on the moving
direction of reticle, secondly tighten another screw by the same mount,