Note: 

The G-TECTA 4G is set to “latch” the alarm 

by default. This means that the instrument will continue 
to alarm until the operator presses the mode button. 

Once you have evacuated the dangerous area, press 
the mode button on the G-TECTA

®

 4G instrument; 

this will reset your G-TECTA

®

 4G to normal operating 

mode. If the G-TECTA

®

 4G continues to alarm, i.e. you 

are still in a dangerous area, pressing the button will 
have no effect. 

 Alarms

The TWA alarm for toxic gases will operate when 
the short-term (15 minute) or long-term (8 hour) 
time weighted average alarm levels are exceeded. 
The G-TECTA

®

 4G will display 

 

and display 

toxic gas levels.

WARNING

 TWA alarms are functional for toxic 

gas instruments only.

WARNING

 The TWA alarms cannot be cleared.

Flammable Over-range Alarm 

If flammable gas levels exceed 100% LEL, G-TECTA

®

 

4G locks into alarm and displays “ ” showing an 
over-range condition. G-TECTA

®

 4G will temporarily 

cut off power to the sensor to prevent burn out, 
and display a progress bar for 200 seconds.

Note:

 The alarm remains active until the instrument 

sensor is in normal operating mode.

When the power timeout is complete, the symbol 
changes to “

” and the mode button should be 

pressed to put the instrument back into normal 
operating mode. The symbol will change to an egg 
timer to indicate that the sensor is warming up.

At the end of the warm up period and assuming the 
instrument is in fresh air, the sensor will revert to a 
numeric display. 

Note:

 If the instrument is still in the hazardous area 

when the sensor is re-set, the sensor will revert to 
an over-range alarm.

The G-TECTA

®

 4G can optionally be configured to 

force the operator to switch the instrument off and 
on to restart. This option is programmable with 
G-TECTA

®

 PC Software.

Note:

 All alarm configurations can be set with 

G-TECTA

®

 PC Software.

WARNING

 If the sensor power is reconnected

when the unit is exposed to an over-range gas
concentration there is a risk of damage to the
pellistor sensor.

WARNING

 Depleted oxygen levels can reduce

the flammable gas reading. If oxygen levels are
below safe breathing limits it should be assumed
that the flammable reading is low.

Instrument Settings

The following instrument settings can be altered 
using G-TECTA

®

 PC Software:

Alarm levels for the sensor: 

Level 1 and level 2 alarms can be set for each 
gas sensor.

Alarm trigger: 

The alarm trigger can be set for rising levels of gas 
(toxic and flammable), or for falling levels of gas 
(oxygen deficiency monitoring).

Содержание G-TECTA 4G

Страница 1: ...G TECTA 4G Multi Gas Detection Instrument Operating Manual UK...

Страница 2: ...ls and information given in both our printed and online publications are accurate at the time of issue full technical specifications are not necessarily included Furthermore The Linde Group policy is...

Страница 3: ...n deficiencies or specific toxic gas accumulations An alarm condition indicates the presence of a potentially life threatening hazard and should be taken very seriously WARNING Ensure only qualified s...

Страница 4: ...s not likely to have ignitable concentrations of flammable gases vapours or liquids present under normal operating conditions Instructions specific for use in non hazardous areas G TECTA 4G with CO2 S...

Страница 5: ...e with the user s normal work function The non slip grip and alligator clip ensure secure handling and wear ability and additional accessories such as shoulder straps and chest harness allow flexible...

Страница 6: ...instrument Note During the start up sequence the next calibration date will be displayed If the calibration due date has expired the G TECTA 4G instrument will display a warning message that calibrati...

Страница 7: ...d Confidence Signals The instrument will display up to four gas readings and each channel will display the gas types units and current readings Screen Icons Always check the gas type being monitored b...

Страница 8: ...t is lowered into a potentially hazardous environment instead of using a sampling hose and also for viewing peak gas exposure at the end of a shift The TWA menu option displays the 8 hour time weighte...

Страница 9: ...bol changes to and the mode button should be pressed to put the instrument back into normal operating mode The symbol will change to an egg timer to indicate that the sensor is warming up At the end o...

Страница 10: ...Weighted Average Alarms TWA alarm on alarm off and peak gas levels during the alarm Zero calibration and gas testing success or failure Flammable sensor alarms on and off The battery condition switch...

Страница 11: ...ry replacement should only be carried out by trained and certified personnel Wearing G TECTA 4G Instruments G TECTA 4G is supplied with a strong and durable alligator clip The alligator clip will secu...

Страница 12: ...ulb at least 10 times using the 2 metre hose supplied Longer lengths of hose and metal probes can be purchased For additional lengths of hose we recommend that the aspirator bulb is squeezed once for...

Страница 13: ...Kit comprises a full gas cylinder and regulator with interconnecting hose a gas test key used to activate Gas Test Mode on the instrument a flow cap to attach to the G TECTA 4G and a vent line The Ga...

Страница 14: ...ress the mode button within 10 seconds to confirm calibration Note If the mode button is not pressed within 10 seconds then the instrument mode will change to Gas Test Mode 5 G TECTA 4G will display a...

Страница 15: ...e instrument must be used for calibrating the other gas sensors G TECTA 4G Instrument with CO2 Sensor When a G TECTA 4G CO2 sensor is zeroed in fresh air it automatically sets the CO2 baseline level t...

Страница 16: ...able on the G TECTA PC software options screen and note the ports available ii Switch on the G TECTA 4G instrument and drop it into the charger cradle iii Check the available communications ports avai...

Страница 17: ...ero not performed from options menu Initiate zero on instrument menu Instrument incorrectly configured Re configure instrument Send to Linde Group service centre G TECTA 4G Product Specification Dimen...

Страница 18: ...TECTA 4G 5 Way Multi charger USA AS4GBCC5AU G TECTA 4G 5 Way Multi charger Australian c w computer interface AS4GBCC5UK G TECTA 4G 5 Way Multi charger UK c w computer interface AS4GBCC5EU G TECTA 4G 5...

Страница 19: ...ediately and return instrument for service Auto Zero failure Instrument zeroed in contaminated atmosphere Switch off instrument and restart in clean air Alarm preventing Auto Zero Instrument zeroed in...

Страница 20: ...able gas reading If oxygen levels are below safe breathing limits it should be assumed that the flammable reading is low G TECTA Electrochemical gas sensors toxic gases or oxygen contain chemicals Ext...

Страница 21: ...oduct conforms with all technical and regulatory requirements of the directives listed below Product Type G TECTA 4G Certification BASEEFA09ATEX0245 IECEx BAS 09 0117 Applicable Directives Electromagn...

Страница 22: ...CULAR PURPOSE THE LINDE GROUP WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY Warranty periods for ins...

Страница 23: ...23 Notes...

Страница 24: ...f going to print Denmark AGA A S Vermlandsgade 55 2300 K benhavn S Tel 45 32 83 66 00 Estonia Eesti AGA AS P rnu mnt 141 11314 Tallinn Tel 372 6504 500 Finland Oy AGA Ab Itsehallintokuja 6 02600 Espoo...

Отзывы: