English 14 English
WARNING
If different
COOL ARC
®
26
is connected, filling procedure
should be started manually.
First follow as shown Figure 33 or 34 and then repeat
steps shown on Figure 15 or 24 to Figure 23 or 32.
Advanced User Interface (U7)
Filling
Cooler
Configuration
Figure 33
Standard User Interface (U0)
Figure 34
Filling procedure failed
Filling the reservoir and water lines first time
If procedure can not be finished at the first time, an
image will be shown on the screen as on Figure 35 or
36.
Figure 35
Figure 36
If during the second attempt filling procedure failed
(Figure 37 or 38) please turn the power source off and
check:
cooling system hoses conditions (bent or damaged)
coolant
level
WAIT
1
2
3
4
5
6
7
8
Procedure failed
Figure 37
Figure 38
After review turn on the power source. Machine will
automatically start with filling procedure again.
WARNING
If the filling procedure fails, contact with the nearest
Technical Service Center or Lincoln Electric. Machine
usage without proper functioning of
COOL ARC
®
26
can
cause thermal damage of water torch or
COOL ARC
®
26
machine.