Note 

• The test leads can add 0.1ȍ to 0.3ȍ of error to resistance measurement. To obtain precision 

readings in low-resistance measurement, that is the range of 400ȍ, short-circuit the input 

terminals beforehand, using the relative function. Push RELǻ button for automatically subtract 

the value from the short-circuited test leads. OL displays when the circuit is open or the resistor 

value is higher than max range. 

 

 

Diode test 

(See fig 3)

 

Use the diode test to check diodes, transistors, and other semiconductor devices. The diode test 

sends a current through the semiconductor junction, and then measures the voltage drop across the 

junction. A good silicon junction drops between 0.5V and 0.8V. 

 

To test a diode out of a circuit, connect as follows: 

1. Insert the red test lead into the HzVȍ terminal and the black test lead into the COM terminal. 

2. Set the rotary switch to diode position (blue). 

3. Push blue button to select diode function. Displays shows diode symbol. 

4. For forward voltage drop readings on any semiconductor component, place the red test lead on 

the component’s anode and  place the black test lead on the component’s cathode. 

The measured value shows on the display. 

 

Continuity test 

(See fig 4) 

To test for continuity, connect as follows: 

1. Insert the red test lead into the HzVȍ terminal and the black test lead into the COM terminal. 

2. Set the rotary switch to continuity position (blue). 

3. Push blue button to select continuity function. Displays shows continuity symbol. 

4. Connect the test leads across with the object being measured. The buzzer sounds if the 

resistance of a circuit under test is less than 70ȍ. 

 

Capacitance measurement 

(See fig 5)

 

1. Insert the red test lead into the HzVȍ terminal and the black test lead into the COM terminal. 

2. Set the rotary switch to capacitance position (blue). 

3. Push blue button to select capacitance function. Displays shows nF symbol. 

4. Connect the test leads to the object being measured. The measured value shows on the display. 

Note 

•When OL displays the capacitor is short-circuit or the selected range is to low. 

•To minimize the measuring error caused by the distributed capacitor, the testing lead should be 

short as possible. It takes longer time when testing a high capacitor value, testing time is around 

15 seconds in 100 µF range. 

• Use the RELǻ function for reduce capacitance stored in the test leads when measure small value 

of capacitance. 

 

Frequency 

( See fig 1 ) 

1. Insert the red test lead into the HzVȍ terminal and the black test lead into the COM terminal. 

2. Set the rotary switch to Hz% position. 

3. Push Hz% button for select measuring frequency in Hz or duty cycle in %. Display shows Hz 

or %.  

3. Connect the test leads across with the object being measured. The measured value shows on the 

display. 

Note. 

When measuring V, µA, mA or A also frequency or duty cycle can be displayed by push 

the Hz% button. Push button by step between frequency, duty cycle or return to previous 

measurement mode. 

Содержание 153220108

Страница 1: ...1 Illustrations Fig 4 Fig 5 Fig 6 Fig 3 Fig 2 Fig 1 ...

Страница 2: ...Tables ...

Страница 3: ...Tables ...

Страница 4: ...Tables ...

Страница 5: ...2 610 English 3 Svenska 8 Norsk 12 Dansk 16 Suomi 21 Deutsch 25 Netherlands 31 Français 36 Italiano 41 Español 46 Português 51 Ελληνικά 56 Polski 62 Eesti 68 Lietuviškai 73 Latviski 79 Pycckий 84 ...

Страница 6: ...nts of non linear signals Reading of Max and Min value Can be connected to PC for transfer data General Specifications Measuring range and accuracy see page 2 Fused Protection for V mA Input Terminal 1 A 250V fast type 6x25 mm Fused Protection for 10 A Input Terminal 10 A 250V fast type 6x25 mm Auto range or manual ranging True RMS for voltage and current measurements Analogue bar graph 61 segment...

Страница 7: ...ehind the finger guards Disconnect circuit power and discharge all high voltage capacitors before testing resistance continuity diodes and current Before measuring current check the Meter fuses and turn off power to circuit before connecting the Meter to the circuit Replace the battery as soon as the battery indicator appears Whit to low battery the Meter might produce false readings that can lead...

Страница 8: ...or himself may be hurt Use proper terminals function and range for the measurement When the testing leads are connect ed to the current terminals do not parallel them across any circuit If the current is more than 5 Amp Measuring time for current should be less than 10 sec and interval time between measurement should be 15 minutes To measure current connect as follows 1 Turn off power to the circu...

Страница 9: ...ch to continuity position blue 3 Push blue button to select continuity function Displays shows continuity symbol 4 Connect the test leads across with the object being measured The buzzer sounds if the resistance of a circuit under test is less than 70ȍ Capacitance measurement See fig 5 1 Insert the red test lead into the HzVȍ terminal and the black test lead into the COM terminal 2 Set the rotary ...

Страница 10: ...uit under test when battery indicator appears on the display 2 Turn the Meter to OFF position 3 Remove the screw and separate the case bottom from the case top 4 Replace the battery with a new 9V battery type 6F22 5 Rejoin the case bottom and case top and reinstall the screw Replace the fuses See figure 6 1 Disconnect the connection between the testing leads and the circuit under test 2 Turn the M...

Отзывы: