Periodic Inspection Sabina
www.liko.com
59
Sabina™
6.0
Instructions for the check points
16 Electrical emergency-lowering device functions
- With the emergency stop out, verify that the electrical emergency-lowering device functions.
17 Charger function
- With the emergency stop out, insert the charger cable into a power outlet (110-240 V AC). The yellow diode light
on the control unit should light up.
Sabina ll:
Make sure that both indicator lamps on the charger are illuminated.
- yellow light during charging
-green light during maintenance charging.
Sabina 200:
When the charger is connected a diode light at the charger indicates;
- yellow light during charging
- green light during maintenance charging
18 Load Test
- Run the lift with specified load* all the way up and down.
Make sure the load is hanging straight down from the
lift hooks, so they don’t bend.
- Run the base in and out with specified load*.
- Listen for peculiar noises and vibrations.
*Test load:
Sabina ll:
Sabina 200:
150kg / 330lbs
120 kg / 265 lbs
19 Documentation
- Instructions/Instruction guide or Quick Reference Guide.
kg / lbs