background image

Laundering: 

See the sling´s product label.

        Medical technical Class I product

© Copyright Liko 

AB 

010-09

Code:

1  Recommended
  Possible
3  Not advisable
4 Inappropriate

Liko HygieneVest has been tested by an accredited test institute and complies with all requirements for MDD Class-I products 

and the harmonized standard EN ISO 10535.

Liko's products are constantly being updated and refined. Liko reserve the right to change aspects of the products without 

prior notice. Contact your local Liko representative for updated information and advice.

 

     

Important!

  

Lifting and transferring a patient always involves a degree 

of risk. Study the instruction guides for both lift systems 

and accessories carefully. A complete understanding of the 

contents of instruction guides is essential.  

Only trained personnel should use the equipment. 

Make certain that the lift accessories are appropriate for the 

lift you are using and for the lifting situation. Exercise caution 

and care when using lift equipment and accessories. 

As a caregiver, you are responsible for the patient’s safety. 

You must be well informed of the patient’s condition and 

ability to manage the lifting situation.

Inspect the sling regularly, especially after laundering. 

Check carefully for wear and damage to seams, fabric, 

straps and strap loops. Never use damaged accessories!  

When in doubt, contact the manufacturer or supplier.

Mini 

0

Slim 350

Standard 450

Wide 670

Recommended combinations of Liko 

HygieneVest and Liko’s sling bars:

Universal 350

Universal 450

Twinbar 670

Teddy HygieneVest, XXS     

50, 55      1 

4

HygieneVest/Teddy HygieneVest, XS 

50, 55 

4

HygieneVest, S      

50, 55 

3

HygieneVest, M                      

50 

 

 

HygieneVest, L   

50 

 

 

HygieneVest, XL   

50 

 

 

 

 

HygieneVest, XXL  

50  

 

 

 

1

HygieneVest, XXXL 

50 

 

1

HygieneVest High Back, M  

55 

 

3

HygieneVest High Back, L 

55 

 

 

3

HygieneVest High Back, XL 

55 

 

 

 

3

The above table summarizes the recommendations for each size. Different sling bar sizes will result in different 

degrees of pressure around the chest. Test various sizes until you find the most appropriate one.

Combinations with Liko HygieneVest and Cross-bar, Cross-bar 450/670 or Side bars is not appropriate.

www.liko.com

Manufacturer: 

Liko AB 

SE-975 9 Luleå

Sweden

[email protected]

Expected Life Time

The product has an expected life time of 1-5 years during normal use. Life time varies depending on fabric, use frequency, 

washing procedure, and weight carried.

Periodic Inspection

The product must be inspected at least once every 6 months. More frequent inspections might be required if the products are 

used or washed more often than normal. See Liko’s protocol Periodic Inspection for the respective product.

Care and Maintenance

Combinations

Combinations of accessories/products other than those recommended by Liko can result in risks for the safety of the patient.
In those cases where another manufacturer recommends Liko’s accessories/products in combination with their or others’ 

accessories/products, and the combinations are not approved by Liko, that manufacturer assumes the responsibility for that 

combination. Liko then refers to the responsible manufacturer’s instruction guides and recommendations concerning usage, 

maintenance, cleaning, checking and inspection of the combination in order to fulfil the EN ISO 10535 requirements.

Other Combinations

Universal 600