LD120
OFSx
(where
OFSx
is the Offset value currently set).
This function is enabled only if
set to “
on
”.
6.2.2 Incremental measurement
Press
P
and
*
key simultaneously to switch from
absolute to incremental measurement.
In incremental mode the decimal point flashes.
Zero setting in incremental mode (see 6.2.1) does
not change the absolute value in the background.
6.2.3 Mm/inch display modes
Mm/inch display mode (and vice versa) can be
changed by pressing
key for 3 seconds.
6.2.4 Setting the offset value
Press
P
and
keys simultaneously to display Offset
1 value (
). Use
and
keys to change value
and save with
*
can be changed only in setup menu.
This function is enable only if
set to “
on
”.
key allows to scroll
OFS1
,
OFS2
and
OFS3
values.
OFS1
= current value +
OFS2
= current value +
OFS3
= current value +
6.2.5 Setting the datum
Press
P
and
keys simultaneously to display datum
. Use
and
keys to change the value
and save with
*
key.
This function is enabled only if
set to “
on
”.
6.3 Setup / Parameter setting
Press
P
key for 3 seconds to enter setup. “
SEtUP
” is
displayed.
Press
key to enter Menu 1 (parameters)
Press
*
key to enter Menu 2 (RS-485 serial I/F)
Press
P
key to access the next Parameter and
Parameter setting.
Press
P
key for 3 s. to exit setup at any moment.
6.3.1 Default parameters
Default parameter values (factory settings) are
highlighted in
BOLD
characters. The unit can be
reset to defaults by pressing
*
key at power-on.
6.3.2 Menu 1: Parameters
rES
[10, 50, 100, 1000, InCH, FrEE] Linear resolution
value expressed in µm (microns).
10
= 10 µm = 0.01 mm
50
= 50 µm = 0.05 mm
100 = 100 µm = 0.1 mm
1000 = 1000 µm = 1 mm
InCH = it sets display mode and measurement unit
to inches (with 3 decimals, e.g. 1.000).
FrEE = it allows to input a free resolution factor,
e.g. for angle display (basis of calculation is
max. resolution 0.01 mm).
*
= save,
P
= next parameter,
P
for 3 s. = exit
FrEE
E.g. We need to display angles ranging between 0°
and 90° with 0.1° resolution on a rotating table
having a circumference of 785.4 mm. Total work
range is 785.4 mm : 4 = 196.35 mm.
= 900 : 19635 = 0.0458
Default value =
0.0001
*
= save,
P
= next parameter,
P
for 3 s. = exit
dEC
[0, 1, 2, 3] Decimal point. Only if
It changes the position of the decimal point.
0 = no decimal point
1 = one decimal (e.g. 1.0)
2
= two decimals (e.g. 1.00)
3 = three decimals (e.g. 1.000)
*
= save,
P
= next parameter,
P
for 3 s. = exit
dir
[uP, dn] Counting direction.
uP
= up (standard direction)
dn = down (inverted direction)
*
= save,
P
= next parameter,
P
for 3 s. = exit
F_rEL
[on, oFF] It enables the incremental measurement
function (by pressing
P
and
*
keys).
on
= function enabled
oFF = function disabled
*
= save,
P
= next parameter,
P
for 3 s. = exit
MAN LD120 I_E 1.6.odt
11
www.lika.it
www.lika.biz