Lighthouse ActiveCount100 Operating Manual
248083444-1 Rev 2
P a g e
37 of 40
“J Smith”
“P Brown”
“Alice Chow”
“Ray Lopez”
. . .
Sampling Run Databases
Sampling Metadata in CSV Files:
ActiveCount100 stores sampling metadata in the following files:
/Active Count 100_<
serial-number
>/acsamples_<
date
>.csv
Where <
date
> is today’s date, e.g., YYYY-MM-DD. ActiveCount100 appends one line of sampling
metadata for each sampling run.
Export sampling metadata to third-party software, such as Laboratory Information Management (LIM)
systems, to complement data from other monitoring systems.
Each sampling database file contains the following comma-delimited fields for each sampling run.
CSV Field
Description
Sample Value
1
Timestamp
Date format – set in Settings menu
US (MM/DD/YYYY)
EU (DD/MM/YYYY)
ISO (YYYY-MM-DD)
Time (local 24-hour time)
HH: MM
7/16/2015 9:48
2
Location
Current location
LOC004
3
User
Current user
Ray Lopez
4
Sample Duration
Total duration of sampling run in minutes
0:10:00
5
Flow Duration
Actual sampling time in minutes
0:10:00
6
Volume
Liters of air sampled
1000
7
Cycles
Constant sampling = 1
Periodic sampling = number of user-specified cycles
1
8
Status
No errors detected = Success
At least one error detected = Error
Success
9
Flow
Flow good = OK
Flow error = Error
OK
10
Service Needed
No service needed = OK
Service needed = Error
OK
11
Calibration
Calibration not yet due = OK
Calibration due = Error
Error
12
Sample Aborted
User did not abort sampling run = OK
Error