Lighthouse REMOTE ActiveCount Installation Guide
4-2
248083413-1 Rev 4
Operating the
RAC System
The Remote ActiveCount can be operated by using the MODBUS
Protocol. The applicable action commands are listed in Table 4-2.
Access RAC
Information
Table 4-3 displays the MODBUS read commands to access RAC
information.
Table 4-2 Action Commands
Value
Register
Action
1
40002
Saves all writable 40002 register values to the
EEPROM.
5
40002
Enable Remote Mode. Locks out the instrument’s
user interface. Can only change instrument
parameters via MODBUS.
11
40002
Instrument Start. Uses defined Initial Delay, Hold
Time, Sample Interval and counting mode. Instrument
executes samples and holds until an Instrument Stop
command is issued. This command will start the
pump.
12
40002
Instrument Stop. Aborts current sample. Stops pump.
Stops data collection.
Table 4-3 Read Commands
Value
Register
Action
50
40001
Read Device Information: Parse through device
information. Verify "VIABLE" in the product name.
1
40003
Read Device Status
2
40005
Read Device Serial Number
74
30001
Get Data Record
72
41001
Get Data Types
72
42001
Get Data Units
72
43001
Get Data Enables
Содержание REMOTE ActiveCount
Страница 1: ......
Страница 2: ...00...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE ActiveCount Installation Guide ii 248083413 1 Rev 4...
Страница 12: ...Lighthouse REMOTE ActiveCount Installation Guide 1 2 248083413 1 Rev 4...
Страница 32: ...Lighthouse REMOTE ActiveCount Installation Guide A 2 248083413 1 Rev 4...
Страница 50: ...Lighthouse REMOTE ActiveCount Installation Guide B 18 248083413 1 Rev 4...
Страница 53: ...00...
Страница 54: ......