Getting Started
248083413-1 Rev 4
3-3
Installation
The RAC Controller and the Remote Switch can be mounted to a
facility wall or customer equipment to allow sampling of critical
locations. Mounting the controller to a wall will allow Grade A
environments to be sampled directly and will allow the remote use of
impactors and isokinetic probes.
Note:
The RAC Controller and Remote Switch must be mounted on a
flat vertical surface.
WARNING:
AC power
must be available within
3 feet of the mounting
location if controller is
wall mounted.
1. Ensure that power has been disconnected from the hub to which the
RAC Controller will be connected.
2. Access the DIP switch and set switches 1 thru 5 to the address
desired. Address 1 is the default address. See Figure 3-1.
Note:
For default RS485
MODBUS operation
switches 1 and 6 must be
ON and switches 7 and 8
must be OFF.
Figure 3-1 DIP Switch
Note:
See Pages 3-5 thru 3-6 for DIP Switch definitions and
addressing.