4.8 SAFETY RELEASE SYSTEM
The safety release system is positioned on the left side of the base unit
(
figure 1
). In the event of an unpredictable machine break down (e.g.
technical defect, power supply failure), this system allows the operator
to manually move down the tracked stairclimber onto the next landing.
By pulling the safety release wire, the electromagnetic brake is
disengaged manually. Due to its own weight (incl. whee weight
of person) the LIFTKAR PTR starts moving downstairs.
ATTENTION:
Before activating the safety release system always press the emergency stop button and remove the key.
Safety first - if safety release system is required, make sure that a second person (helper) is present, who
activates the safety release system.
By pulling the safety release wire (helper) and pressing the handle unit downwards (operator), the
device starts moving.
The device can be stopped anytime by releasing the safety release wire. Only move the stairclimber onto the
next landing.
If the total weight (tracked stairclimber and passenger) is low, the operator will need to support travel down
the stairs by pushing in the direction of travel.
After usage of the safety release system your LIFTKAR PTR must undergo a complete safety check by an
authorised SANO partner.
5 CLEANING AND DISPOSAL
5.1 CLEANING AND DISINFECTION
Our motto is:
A CLEAN MACHINE MOVES MORE SMOOTHLY.
Cleaning with a standard household detergent is quite sufficient.
Never use a high-pressure cleaner or similar.
If required, the surfaces can be cleaned with a commercially available alcohol-based equipment disinfection
agent (such as Sagrotan®).