4
Technical Information and Specifications
ELECTRICAL:
This System is a Class I, Type B Applied Part.
This ME equipment conforms to AAMI Std ES60601-1 & IEC Std
60601-1 and is certified to CSA Std C22.2 No. 60601-1.
This ME equipment is tested to
IEC 60601-1-2:2014, Edition 4.0
ELECTRICAL RATINGS for Models 120G5 and 240G7
MODEL 120G5:
110-120 V~ 60 Hz unit is rated for 2A (U.S.,
Canada)
MODEL 240G7:
220-240 V~ 50/60 Hz unit is rated for 1A (EU, Asia)
POWER CORDS
This system comes with a ten-foot-long, hospital grade power cord
manufactured by Volex or equivalent (UL, CSA Listed) for the
110-120 V~ 60 Hz unit.
This system comes with an eight-foot-long, hospital grade power
cord manufactured by Volex or equivalent (UL, CSA Listed) for the
220-240 V~ 50/60 Hz unit.
FILTERS
Pre-Filter: Rated per filter manufacturer's instructions
UV Sediment: Rated at 16,000 gallons or 4 months
NORMAL USE
For this system in normal use, the patient is positioned on basin
cabinet with nozzle inserted for duration of procedure (30-40
minutes). The operator is in the room monitoring patient. There is
no change of setting of a control. The operator does not touch the
patient while touching the viewing tube bulb holder or the
UV harness assembly. In the event of a power outage to the system
or any other performance change or change in the operating
environment, the operator should have the patient slide off the rectal
nozzle and stop the session. The procedure may resume once
normal operating conditions have been restored.
APPLIED PARTS
Colon cleansing nozzle (rectal nozzle), backrest, basin cabinet