Installation Guide  

LC-15-100-002.F (rev 01.01.20)

 

 

Angel of Water

®

 CM-1 Series (120G5/240G7)

 

HyGIeaCare

® Prep System (120G5/240G7) 

 

 

 

 

ETL CLASSIFIED

 

Lifestream Purification Systems, LLC 

7303 Burleson Rd., Suite 801

 

Austin, TX  78744  USA 

TEL +1.512.707.8383

 

Toll Free U.S. 877.564.3185

 

lifestream.systems

  

 

 

 

 

 

4001754

 

2797 

Содержание 120G5

Страница 1: ...0 Installation Guide LC 15 100 002 F rev 01 01 20 Angel of Water CM 1 Series 120G5 240G7 HyGIeaCare Prep System 120G5 240G7 ETL CLASSIFIED Lifestream Purification Systems LLC 7303 Burleson Rd Suite 80...

Страница 2: ...n Guide is proprietary to Lifestream Purification Systems LLC and is intended solely for the use of purchasers and operators of the medical device s herein described Reproduction in any manner is proh...

Страница 3: ...TABLE OF CONTENTS Technical Information and Specifications 6 WARNINGS CAUTIONS 8 Steps to Installation 12 Appendix A Electromagnetic Compatibility Information 30...

Страница 4: ......

Страница 5: ...S NOTE Check ALL parts for shipping damage If shipping damage is noted DO NOT USE Contact carrier or manufacturer for further instructions NOTICE The information contained in this document is subject...

Страница 6: ...s device to sale by or on the order of a physician ATTENTION Uniquement sur ordonnance Dangerous voltage Caution Warning dangerous voltage General warning sign Manufacturer YYYY Date of manufacture UV...

Страница 7: ...used for fluid storage however residual risk has not been evaluated for children or for pregnant or nursing women Authorized representative in the European Community Lower limit of temperature Upper l...

Страница 8: ...Volex or equivalent UL CSA Listed for the 220 240 V 50 60 Hz unit FILTERS Pre Filter Rated per filter manufacturer s instructions UV Sediment Rated at 16 000 gallons or 4 months NORMAL USE For this sy...

Страница 9: ...rature safe range 37 C 99 F 39 8 C 103 F PRESSURE FLOW RATE AND WATER TEMPERATURE TO THE ME EQUIPMENT FROM MUNICIPAL WATER SOURCE Pressure of incoming water not to exceed 100 p s i NOTE Municipal wate...

Страница 10: ...p d une terre de protection WARNING Connecting electrical equipment to MSO effectively leads to creating an ME SYSTEM and can result in a reduced level of safety Responsible organization is referred t...

Страница 11: ...R SERVOIR WARNING Do not use where water is microbiologically unsafe or with water of unknown quality without adequate disinfection before and after the unit AVERTISSEMENT Ne pas utiliser lorsque l e...

Страница 12: ...tem in an MRI environment or in the vicinity of high frequency surgical diathermy equipment defibrillators or shortwave therapy equipment Electromagnetic interference could disrupt the operation of th...

Страница 13: ...0 75 non condensing 0 75 non condensing 0 95 non condensing 700 hPa to 1060 hPa 700 hPa to 1060 hPa a 3000 m Les conditions environnementales Temp rature fonctionnement stockage transport Humidit rel...

Страница 14: ...to be placed in either Range A or B pictured You may choose A or B depending upon procedure room constraints The Exhaust Vent Opening 1 1 2 3 8 cm pipe See Step 11 needs to be placed in the ceiling ab...

Страница 15: ...cm from Back Corner 72 182 9 cm Basin Cabinet 17 12 43 2 cm 30 5 cm Tower Cabinet Door When Opened Total System Width Approx 32 5 82 6 cm Approx 101 256 5 cm Total System Length with door open At leas...

Страница 16: ...CAL CODE REQUIREMENTS FOR YOUR HEALTH CARE FACILITY 1 Side Wall Placement of Electrical Outlet must be within 18 45 cm of the floor If side wall is used for outlet make placement between 12 and 40 inc...

Страница 17: ...e with basin cabinet placement later Figure 3 1 2 The system requires both one hot and one cold line as suggested bring 1 inch water lines to the procedure room for optimal water pressure To the 1 inc...

Страница 18: ...tower male fitting and mock up any additional rough plumbing fittings and pipe lengths necessary before gluing Confirm drain connection is correct for proper drainage Then glue components into place...

Страница 19: ...the 2 electrical cords into tower cabinet rectangular opening to plug into each of their respective tower large junction box outlets later See Figure 5 2 3 Align basin cabinet drain fitting to tower c...

Страница 20: ...16 Figure 5 3 Basin Tower...

Страница 21: ...inside tower wall near bottom of cabinet See Figure 6 1 2 Now place Cup and Drain Assembly supplied by Lifestream on floor See Figure 6 1 Slide into place and connect both female unions to male threa...

Страница 22: ...18 3 Make sure Tank Exit is aligned over Cup and Drain Assembly This is an Air Gap which keeps the tank above sanitary and the drain below separate See Figure 6 2 Figure 6 2 Air Gap...

Страница 23: ...e correctly or the corresponding function will not work The cords are color matched Although the female end of the power cord is shown in Figure 7 1 as plugged in to its inlet for placement DO NOT plu...

Страница 24: ...in 1 2 inch 1 3 cm connecting pipes supplied by Lifestream with fittings as shown Figure 8 1 2 Connect those pipes and gaskets by hand tightening them securely to front of basin and front of tower as...

Страница 25: ...ose connectors not included CONSULT YOUR LICENSED QUALIFIED PLUMBER NOTE HOT Line is RED left COLD Line is BLUE right Temperature of hot water from hot water supply to equipment 10 valve should not ex...

Страница 26: ...tower cabinet to the basin cabinet 2 Tower drain assembly installed inside tower cabinet 3 All electrical cord connectors between basin and tower cabinets plugged in 4 Both basin to tower connectors i...

Страница 27: ...lumbing Connect a 1 inch 3 8 cm draft pipe not included to the exhaust port fitting at the top of the tower and vent either through side wall or through the ceiling to the outside Manufacturer s Recom...

Страница 28: ...ndow in the pre drilled holes Figure 12 1 Chain Hinge 2 Eyelet 1 Screw small eyelet provided into top pre drilled hole 2 Remove and then connect screw on lower hole on front of the tower cabinet to in...

Страница 29: ...to Electrical Source 1 Plug the female end of the power cord into the Junction Mains Box inlet Refer to Figure 7 1 NOTE Lifestream supplies a 110 120 V or 220 240 V power cord that is country specific...

Страница 30: ...acuum Breakers described on P 15 of the User s Guide by doing the following A LOCATE the two shutoffs on each of the vacuum breakers one below the main body of the valve inlet one above the main body...

Страница 31: ...at the bottom of the tower is sufficiently tightened to prevent leaking while Viewing Tube Sprayer is ON CLOSE sprayer 8 OPEN volume control for Basin Sprayer and CHECK it by spraying water into the...

Страница 32: ...Junction Mains Box in the tower is turned ON refer to Figure 7 1 Turn on LIGHTS Switch on Tower Switch Panel to make sure that tower and basin lights come on Figure 14 4 Tower Switch Panel Remote Swit...

Страница 33: ...nstallation Guide and sanitized according to the User s Guide it is ready for use For questions regarding installation please contact Lifestream Lifestream Purification Systems LLC 7303 Burleson Rd Su...

Страница 34: ...es other than those specified with the exception of transducers and cables sold by the manufacturer of this device as replacement parts for internal components may result in in creased Emissions or de...

Страница 35: ...Group 1 The system uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR...

Страница 36: ...to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power...

Страница 37: ...ulated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power...

Страница 38: ...ulated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power...

Страница 39: ...ic interference by maintaining a minimum distance be tween portable and mobile RF communications equipment transmitters and the system as recommended below accord ing to the maximum output power of th...