background image

Содержание UA-767T

Страница 1: ...UA 767T Blood Pressure Monitor Talking One Step Auto Inflation ...

Страница 2: ... that daily self monitoring of blood pressure is an important way individuals can contribute to maintaining their cardiovascular health and preventing the serious consequences of hypertension left undiagnosed and untreated LifeSource has been manufacturing quality healthcare products for more than 20 years Rest assured that we are committed to providing you and your family with monitoring devices ...

Страница 3: ... BEFORE YOU START 5 SELECTING THE CORRECT CUFF 6 TAKING YOUR BLOOD PRESSURE 7 10 CHECK ADJUST EXHAUST VELOCITY 11 12 ABOUT BLOOD PRESSURE 13 15 BLOOD PRESSURE RECORD 16 IMPORTANT INFORMATION 17 SPECIFICATIONS 18 CONTACT INFORMATION 19 LifeSource Health Line Toll Free 1 888 726 9966 2 ...

Страница 4: ...OR 1 increments of between 2 mmHg and 5 mmHg r Appears if pressure Try measurement again 2 value is unstable due and remain very still to movement during ERROR 2 reading r Appears when pressure Make sure cuff tube is rr value did not increase properly connected to 3 during cuff inflation monitor cuff is securely ERROR 3 1 cuffis not fastened fastened and exhaust rate 2 air connector is back is bet...

Страница 5: ...S LOW BATTERY DISPLAY tg I __ PRESSURE l g SWITCH t DIASTOLIC DISPLAY START BUTTON VOLUME CONTROL PULSE DISPLAY AIR CONNECTOR CUFF r T BATTERY COMPARTMENT ADJUSTMENT SCREW 1 HEADPHONE SOCKET TUBE BLACK PROTECTION CAP ON ADJUSTMENT SCREW LifeSource Health Line Toll Free 1 888 726 9966 o 4 ...

Страница 6: ... and contracts against the cuff with each heartbeat BEFORE YOU START D You must install 4 type AA 1 5 volt batteries and attach the cuff to the monitor before using it To install batteries or replace them if the Low Battery symbol appears on display proceed as follows LifeSource Health Line Toll Free 1 888 726 9966 5 1 Remove battery compartment cover by gently pushing down on arrow and sliding co...

Страница 7: ...the circumference of your upper arm at the midpoint between shoulder and elbow ARM SIZE ECOMMENDED REPLACEMENT UFF SIZE CUFF MODEL 5 1 7 9 13 20cm Small UA 279 7 5 12 2 19 31cm Medium UA 280 11 8 17 7 30 45cm Large UA 281 AIR CONNECTOR TUIi MONITOR To attach cuff insert air connector at end of cuff tube into air socket on left side of monitor Should you need to change the cuff size the preset exha...

Страница 8: ...able to insert two fingers between the cuff and your arm Tips for Blood Pressure Monitoring IBF Relax for about 5 to 10 minutes before measurement IBF Remove constricting clothing and place cuff on bare arm IBF Unless your physician recommends otherwise use left arm to measure pressure Now you are eady Follow these simple steps 1 Sit comfortably with your left arm resting on a flat surface so that...

Страница 9: ...essure at the current setting PRESSURE 240 G 210 180 150 Usual Systolic Set at Up to 120 150 121 150 180 151 180 210 181 210 240 211 see NOTE NOTE If your systolic reading is greater than 210 please press and hold the START button until the pressure switch reaches 30 mmHg higher than your systolic reading The maximum cuH pressure is set at 320 mmHg If it reaches 320 mmHg the auto exhaust will be t...

Страница 10: ...d diastolic pressure readings and pulse rate are then displayed and announced before the cuff deflates fLU N 6q The announcement repeats four times before the monitor automatically shuts off You may also stop the announcement and turn off the monitor at any time by pressing the START button To increase the volulne of the announcement move the VOLUME CONTROL switch see diagram page 4 to the right T...

Страница 11: ...r time NOTE If you wish to take your blood pressure again relax and wait 5 10 minutes to enable the flow of blood in the arm to return to normal There is no need to wait if someone else wishes to use the monitor USING THE HEADPHONES Your Talking Blood Pressure Monitor may be used with headphones The HEADPHONE SOCKET is located on the lower right edge of the monitor Insert jack into the socket Use ...

Страница 12: ...m size If using a cuH other than the one shipped with the unit you may need to adiust the exhaust velocity rate There is no need to adiust exhaust velocity when changing between a medium cuH and a large cuH with a valve To check the exhaust velocity rate proceed as follows 1 Put on the arm cuff 2 Press the START button The unit will begin taking a reading 3 Watch the display screen as the cuff def...

Страница 13: ...thin plastic disk proceed to step 2 REMOVE THE COVER FROM THE ADJUSTMENT SCREW 2 Using a small screwdriver rotate adjustment screw about 45 degrees clockwise to raise the rate or counterclockwise to lower the rate 3 Repeat the test until the exhaust velocity is between 2 and 5 mmHg NOTE Once the monitor is adiusted to a new cuH size further adiust ments should not be required unless cuH size is ch...

Страница 14: ...king your blood pressure repeatedly without waiting an interval of 5 to 10 minutes between readings or without raising your arm to allow blood to flow back to the heart can also affect it In addition to these factors diet beverages containing caffeine or alcohol certain medications emotional stress and even tight fitting clothes can make a difference in the readings What Causes Variations In Blood...

Страница 15: ...o stroke heart attack or other illness if left untreated Referred to as a silent killer because it does not always produce symptoms that alert you to the problem hypertension is treatable when diagnosed early Can Hypertension Be Controlled In many individuals hypertension can be controlled by altering lifestyle and minimizing stress and by appropriate medication prescribed and monitored by your do...

Страница 16: ...ings In any case self measurement at home supplements the doctor s readings and provides a more accurate complete blood pressure history In addition clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home How Do I Record My Blood Pressure Blood pressure readings are typically r...

Страница 17: ...B L 0 0 D P RESSURE RECORD Name Age Weight DATE AM SYS DIA PULSE PM SYS DIA PULSE 1 14 9 30 132 9cS YJ69 6 30 12CS CS7 YJ63 E LifeSource Health Line Toll Free 1 888 726 9966 o 16 ...

Страница 18: ...t to extremes in temperature or humidity or to direct sunlight shock and dust LifeSource guarantees the accuracy of this device only when it is stored and used within the temperature and humidity ranges noted on page 18 Clean the monitor and cuff with a dry soft cloth or a cloth dampened with water and a mild detergent Never use alcohol benzene thinner or other harsh chemicals to clean monitor or ...

Страница 19: ...proximately 4 months with one daily measurement Operating environment 50 F to 104 F 10 C to 40 C Less than 85 relative humidity Storage environment 4 F to 158 F 20 C to 70 C Less than 95 relative humidity Dimensions Length 4 2 11 106 mm Width 6 4 11 063 mm Height 2 8 11 70 mm Weight 11 3 oz 321 g without batteries Blood pressure measurements determined by the UA 767T are equivalent to those obtain...

Страница 20: ...s LifeSource product is covered by a Lifetime Warranty See warranty card for details For more information regarding use care or servicing of your Blood Pressure Monitor contact LifeSource A division of A D Engineering Inc 1555 McCandless Drive Milpitas CA 95035 LifeSource Health Line Toll Free 1 888 726 9966 www LifeSourceOnHne com For Canada Residents please contact Auto Control Medical 206A boul...

Страница 21: ...raille or an audio manual for Model 767T please call LifeSource Health Line at 1 888 726 9966 LifeSource A division of A D Engineering Inc 1555 McCandless Drive Milpitas CA 95035 www LifeSourceOnline com UA 767T Printed in Japan 12 99 ...

Отзывы: