LifePort Kidney Transporter Operator’s Manual 755-00002 Rev R
23
Remove Kidney From LifePort Kidney Transporter
The procedure for removing the kidney from LifePort Kidney Transporter is detailed below. This procedure may
be modified as necessary.
WARNING:
Where noted, perform the following procedure on an aseptic field using standard aseptic
technique.
1. Unlatch and remove LifePort Kidney Transporter Cover.
2. Remove the Outer Perfusion Circuit Lid and place upside-down on a table where it will be undisturbed.
3.
Using standard aseptic technique,
carefully place LifePort Kidney Transporter Disposable Sterile Drape
onto LifePort Kidney Transporter, lining up the drape gasket with the Organ Chamber. Ensure the arrow on
the orientation guide is pointing toward the Pump Deck.
4.
Using standard aseptic technique,
unfold Sterile Drape in the following order:
right
,
left
,
front
, and
back
.
The drape gasket should fit securely around the Organ Chamber with the tabs snapping into place under
the lid latches.
5.
Using standard aseptic technique,
unlatch and remove the Inner Perfusion Circuit Lid and place it
facedown onto the sterile field.
6. Press the
STOP
button.
7.
Using standard aseptic technique,
unscrew or cut the Infuse Line.
8.
Using standard aseptic technique,
carry the
Kidney Cradle, with the cannulated kidney, to the sterile field.
9.
Using standard aseptic technique,
unhook the Mesh Organ Restraint.
10.
Using standard aseptic technique,
unstrap, open, and remove the cannula.
11. Once the kidney has been removed from LifePort Kidney Transporter and accepted by the transplant
surgeon, proceed to
Clean and Disinfect After Use
.
Clean and Disinfect After Use
Once the kidney has been removed from LifePort Kidney Transporter Disposable Perfusion Circuit, replace
both Perfusion Circuit Lids and power off LifePort Kidney Transporter.
All LifePort Kidney Transporter Disposables and perfusate are single-use devices and should go into medical
waste disposal.
WARNING:
Use Universal Precautions when performing
perfusate and equipment cleanup to prevent
possible contact with bloodborne pathogens.
LifePort Kidney Transporter does not come in contact with the donor organ. The donor organ should always be
within the sterile field provided by
LifePort Kidney Transporter Disposable Perfusion Circuit and Sterile Drape.
LifePort Kidney Transporter must be thoroughly cleaned and disinfected after each use. Prior to cleaning and
disinfecting, assemble the following agents and supplies:
• 70% isopropanol (solution, wipes, or swabs)
• Hospital grade pre-moistened germicidal wipes (Super Sani-Cloth
®
, CaviWipes
TM
)
• Soft, lint-free cloths
• Water
WARNING:
Do not clean LifePort Kidney Transporter while plugged into AC mains.
WARNING:
Do not use cleaning solutions containing acetone, ammonia, benzene, xylene, or similar
solvents. Do not use abrasive cleaning tools or pressurized spraying devices. Do not clean or disinfect
by autoclave or sterilize with EO gas. Doing so will void the warranty.