LifeAssays Magnia Reader Скачать руководство пользователя страница 8

Page 8 of 8

7.5  Storing or transporting Reader

To make sure the Reader operates correctly, the following conditions for 

the storing and transportation must be met:

•  Store the Reader and the batteries in dry conditions in the temperature 

between -20ºC – +50 ºC.

•  Store the batteries fully recharged, detached from the Reader.
•  Remove the carrier when transporting and / or storing the Reader to 

avoid damage to the Reader.

•  Reattach the piece of cardboard back into the swing bed to protect it 

during transportation and / storing.

8 Troubleshooting

8.1  Warning and error messages

The Reader display shows warning and error messages.

Warning messages inform about a problem or a failure, but it is possible 

to continue working with the Reader after receiving one.

Warning messages

Explanation

USE RECHARGER

The Reader battery is running out of 

power.

You should recharge the battery as soon 

as possible. You can continue working 

while the battery is recharging.

NOT CALIBRATED

Please contact Lifeassays AB.

TEMPERATURE HIGH

High temperature of the Reader may affect 

the measurement results.

This warning will disappear when suitable 

operating temperature has been reached.

TEMPERATURE LOW

Low temperature of the Reader may affect 

the measurement results.

This warning will disappear when suitable 

operating temperature has been reached.

NOISE LEVEL HIGH

Please contact Lifeassays AB.

REPLACE BATTERY

The Reader battery is reaching the end of 

its lifespan.

The battery needs to be replaced. New 

batteries can be ordered from Lifeassays AB.

Error messages inform about a problem or failure and require actions 

before you can continue working with the Reader, for example pressing 

the POWER button.

NOTE

if pressing POWER button does not solve the 

situation, contact Lifeassays AB.

Warning messages

Explanation

BATTERY 

VOLTAGE LOW

The Reader battery is running out of 

power.

Connect the Reader to the recharger to 

be able to carry out the measurement 

sequence. You can continue working while 

the battery is recharging.

RECEIVER FAILURE

Electronics failure in the receiver.

Press POWER button to return to the main 

display view.

TRANSMITTER FAILURE

Electronics failure in the transmitter.

Press POWER button to return to the main 

display view.

POSITIONING FAILED

The position of the carrier is incorrect.

Press POWER button to return to the main 

display view.

MOTOR 

CURRENT HIGH

The carrier could not be lifted.

Press POWER button to return to the main 

display view.

9  Disposing of Reader

When the Reader has reached the end of its lifespan, it must be disposed 

of with other electronic waste. The battery must be removed and 

disposed of according to the national and / or regional legislations.

Magnia is a registered trademark of Magnasense Technologies Oy. 

Latest update of the User Manual 27.09.2017

Manufactured for LifeAssays AB by 

Magnasense Technologies Oy

Rajatorpantie 41C

FI-01640 Vantaa

FINLAND

LifeAssays AB, IDEON Science Park, SE-223 70 Lund, Sweden

Phone: +46 46 286 54 00, Email: [email protected], Web: www.lifeassays.com

E-25-0025-01 

Содержание Magnia Reader

Страница 1: ...d safety instructions specified in this manual Lifeassays AB does not accept liability for any damage to people or equipment due to the misuse of the Reader 2 1 Contact information and spare parts LifeAssays AB Sölvegatan 43A 22370 Lund SWEDEN Phone 46 46 286 54 00 Email info lifeassays com Web www lifeassays com 3 Contents of delivery The following items are included in the delivery 1 User Manual...

Страница 2: ...tus of the battery recharging When the battery is being recharged the light blinks When the light stays on the battery is fully recharged No function Reserved for future updates No function Reserved for future updates No function Reserved for future updates No function Reserved for future updates No function Reserved for future updates 3 Test cartridge A test cartridge contains the test that is be...

Страница 3: ... last row gives instructions on what to do to proceed with the selected task for example Insert Test Warning messages are shown on the second last row if there are any and the row is not otherwise used Normally the second last row indicates the battery status The warning messages are prioritized and the most critical is always shown During the testing sequence the display shows the name of the tes...

Страница 4: ...r transportation its battery must be connected and recharged for recharging see chapter 5 3 When taking first time the reader into use the battery is already located inside the Reader but the battery connection plug is detached for shipping To connect the battery 1 Turn the Reader upside down The battery hatch is in the bottom of the Reader 2 Remove the screw of the battery hatch 3 Turn the Reader...

Страница 5: ...ut field of the software that is used for recording information 3 Open the aluminum test pouch having one LifeAssays test device and one disposable pipette inside Insert the LifeAssays test device into the Reader so that the small arrow on the end of the test device is visible and pointing towards you The sample well marked with S will remain outside the Reader The Reader prompts Use Calibration K...

Страница 6: ...pected or Reader does not meet user defined system suitability test 7 1 1 Manual calibration check To check calibration manually 1 Turn on the Reader 2 Insert the Check Calibration card in the Reader 3 Insert the calibration cartridge 4 Press the Measure button 5 After the results are ready record them 6 Compare the results with the allowed values printed on the Calibration cartridge 7 Repeat the ...

Страница 7: ...version on the display for a few seconds and then returns to the main display 3 Remove the Software update memory card from the Reader The Reader is now ready to use NOTE Software version of the Reader is shown on the display always when the Reader is turned on 7 4 Replacing battery If the Reader gives a warning Replace battery it means the battery is closing the end of its lifespan and needs to b...

Страница 8: ...aching the end of its lifespan The battery needs to be replaced New batteries can be ordered from Lifeassays AB Error messages inform about a problem or failure and require actions before you can continue working with the Reader for example pressing the POWER button NOTE if pressing POWER button does not solve the situation contact Lifeassays AB Warning messages Explanation BATTERY VOLTAGE LOW The...

Отзывы: