LifeAssays Magnia Reader Скачать руководство пользователя страница 1

LifeAssays AB, IDEON Science Park, SE-223 70 Lund, Sweden

Phone: +46 46 286 54 00, Email: [email protected], Web: www.lifeassays.com

Table of contents

1 Introduction

The purpose of this manual is to provide instructions on how to safely 

and correctly use and service Magnia Reader (also “Reader” in this 

document).

Here are the symbols that are used throughout this documentation:

 

WARNING

Information presented in this box warns about 

serious danger or damage to people and/or 

to equipment that may also lead to unreliable 

results. The instructions offer a safe way of 

proceeding

 

CAUTION

Information presented in this box warns about 

possible damage to equipment and may also 

lead to unreliable results. The instructions offer a 

safe way of proceeding.

NOTE

Information presented in this box needs to be 

especially carefully considered.

2  Magnetic Immunoassay Reader

The Reader is used for analyzing the magnetic particle content in various 

types of immunodiagnostic lateral flow tests. It offers superior detection 

limit for particles and a wide dynamic range and sensitiveness for 

measuring analytes under investigation.

The Reader supports analyzing several different types of test cartridges. 

New types of tests can be introduced without the need for alterations in 

the Reader.

NOTE

The Reader may only be used for the purpose 

defined in this chapter, following the user and 

safety instructions specified in this manual.

Lifeassays AB does not accept liability for any 

damage to people or equipment due to the 

misuse of the Reader.

2.1  Contact information and spare parts

LifeAssays AB

Sölvegatan 43A

22370 Lund

SWEDEN

Phone:+46 46 286 54 00

Email: [email protected]

Web: www.lifeassays.com

3  Contents of delivery

The following items are included in the delivery:

•  1 User Manual E-25-0025
•  1 Magnia Reader art no. 50101-10
•  1 USB cable art no. 50201-13
•  1 carrier art no. 50201-14
•  1 battery recharger art no. 50201-16
•  1 “Check calibration” card for manual calibration check  

art no. 50201-16

•  2 calibration cartridges for manual calibration check 

In case some of these listed items are missing or damaged during 

transportation, please contact Lifeassays AB.

 

WARNING

Do not use damaged equipment. It may cause 

serious damage or danger to people and/or 

equipment.

Magnetic Immunoassay  

Reader – Operator Manual

Page 1 of 8

E-25-0025-01 

1 Introduction ......................................................................1
2  Magnetic Immunoassay Reader ....................................1

2.1  Contact information and spare parts ......................1

3  Contents of delivery .........................................................1
4  Technical information ......................................................2

4.1  The display ...............................................................3
4.2 Carrier .......................................................................4

5. Setting up Reader...............................................................4

5.1  Connecting battery ..................................................4
5.2  Recharging battery ..................................................4
5.3  Activating Reader .....................................................5
5.4  Replacing carrier ......................................................5

6  Operating Reader ...............................................................5

6.1 Measuring .................................................................5

7  Maintenance ........................................................................6

7.1  Checking calibration ................................................6

7.1.1 Manual calibration check .................................6
7.1.2 Cleaning sensor head ......................................6

7.2  Cleaning Reader .......................................................7
7.3  Updating software ....................................................7
7.4  Replacing battery .....................................................7
7.5  Storing or transporting Reader ................................8

8 Troubleshooting ...............................................................8

8.1  Warning and error messages ...................................8

9  Disposing of Reader ........................................................8

Содержание Magnia Reader

Страница 1: ...d safety instructions specified in this manual Lifeassays AB does not accept liability for any damage to people or equipment due to the misuse of the Reader 2 1 Contact information and spare parts LifeAssays AB Sölvegatan 43A 22370 Lund SWEDEN Phone 46 46 286 54 00 Email info lifeassays com Web www lifeassays com 3 Contents of delivery The following items are included in the delivery 1 User Manual...

Страница 2: ...tus of the battery recharging When the battery is being recharged the light blinks When the light stays on the battery is fully recharged No function Reserved for future updates No function Reserved for future updates No function Reserved for future updates No function Reserved for future updates No function Reserved for future updates 3 Test cartridge A test cartridge contains the test that is be...

Страница 3: ... last row gives instructions on what to do to proceed with the selected task for example Insert Test Warning messages are shown on the second last row if there are any and the row is not otherwise used Normally the second last row indicates the battery status The warning messages are prioritized and the most critical is always shown During the testing sequence the display shows the name of the tes...

Страница 4: ...r transportation its battery must be connected and recharged for recharging see chapter 5 3 When taking first time the reader into use the battery is already located inside the Reader but the battery connection plug is detached for shipping To connect the battery 1 Turn the Reader upside down The battery hatch is in the bottom of the Reader 2 Remove the screw of the battery hatch 3 Turn the Reader...

Страница 5: ...ut field of the software that is used for recording information 3 Open the aluminum test pouch having one LifeAssays test device and one disposable pipette inside Insert the LifeAssays test device into the Reader so that the small arrow on the end of the test device is visible and pointing towards you The sample well marked with S will remain outside the Reader The Reader prompts Use Calibration K...

Страница 6: ...pected or Reader does not meet user defined system suitability test 7 1 1 Manual calibration check To check calibration manually 1 Turn on the Reader 2 Insert the Check Calibration card in the Reader 3 Insert the calibration cartridge 4 Press the Measure button 5 After the results are ready record them 6 Compare the results with the allowed values printed on the Calibration cartridge 7 Repeat the ...

Страница 7: ...version on the display for a few seconds and then returns to the main display 3 Remove the Software update memory card from the Reader The Reader is now ready to use NOTE Software version of the Reader is shown on the display always when the Reader is turned on 7 4 Replacing battery If the Reader gives a warning Replace battery it means the battery is closing the end of its lifespan and needs to b...

Страница 8: ...aching the end of its lifespan The battery needs to be replaced New batteries can be ordered from Lifeassays AB Error messages inform about a problem or failure and require actions before you can continue working with the Reader for example pressing the POWER button NOTE if pressing POWER button does not solve the situation contact Lifeassays AB Warning messages Explanation BATTERY VOLTAGE LOW The...