WHAT DISPLAY SYMBOLS MEAN
?
Display
Symbol
Recommended
Action
Measurement is in progress, remain
as still as possible.
Take measurement again and con-
sult with your physician.
Replace all batteries with new ones
when the symbol blinks. Alkaline
batteries recommended.
Try the measurement again. Remain
very still during the measurement.
Fasten the cuff correctly, and try the
measurement again.
Check for air leaks along the tube
and around the
air socket.
Refasten the cuff and retake meas-
urement.
Make sure tube is properly connect-
ed to cuff and monitor.
Try the measurement again. Remain
very still during the measurement.
Measurement is in progress, remain
as still as possible.
Refer to the section “About Pressure
Rating Indicator” for further expla-
nation.
Press exhaust button to release air.
Inflate cuff 30 to 40 mmHg above
your normal systolic pressure.
Condition/Cause
The symbol shown while measure-
ment is in progress. It blinks while
detecting the pulse.
An irregular heartbeat or body move-
ment may have occurred. Refer to
page E-9 for more information on
irregular heartbeats.
Previous measurement stored in
MEMORY.
The battery power indicator during
measurement.
The batteries are low when it blinks.
Unstable blood pressure due to
excessive body movement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during inflation.
The cuff is not fastened correctly.
There is an air leak in the cuff or
monitor.
The pulse is not detected correctly.
Cuff Inflation Meter
Pressure Rating Indicator™
Appears on display prior to and dur-
ing measurement. Blinks if there is air
in cuff at power up.
Cuff needs to be inflated for proper
measurement.
Battery Full
Battery Low
PULSE DISPLAY ERROR
?