In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
SPECIFICATIONS
Model
. . . . . . . . . . . . . . . . . . . . . UA-705
Type
. . . . . . . . . . . . . . . . . . . . . . Oscillometric
Display
. . . . . . . . . . . . . . . . . . . Digital character height; Pressure (15.0
mm)/pulse (10.0 mm) display
simultaneously
Memory
. . . . . . . . . . . . . . . . . . . 30 readings
Measurement range
. . . . . . . Pressure: 20 mmHg to 280 mmHg
Pulse: 40 pulses to 200
pulses/minute
Accuracy
. . . . . . . . . . . . . . . . . . Pressure:
±
3 mmHg or
±
2%,
whichever is greater Pulse:
±
5%
Pressurization
. . . . . . . . . . . . Manually by inflator bulb
Depressurization
. . . . . . . . . Constant-air release-valve system
Deflation
. . . . . . . . . . . . . . . . . . Hand-held push button
Power source
. . . . . . . . . . . . . 1 type "AA" (1.5 volt) battery (included)
Battery life
. . . . . . . . . . . . . . . . Approximately 2,000 measurements
Operating environment
. . . 50ºF to 104ºF (10ºC to 40ºC)
Less than 85% relative humidity
Storage environment
. . . . . . -4ºF to 140ºF (-10ºC to 60ºC)
Less than 95% relative humidity
Dimensions
. . . . . . . . . . . . . . . Width: 3.2" (81 mm)
Depth: 4.2" (105mm)
Height: 2.2" (54 mm)
Weight
. . . . . . . . . . . . . . . . . . . . 4.3 oz. (120 g) without battery
Blood pressure measurements determined by the UA-705 are equiva-
lent to those obtained by a trained observer using the cuff/stetho-
scope auscultation method within the limits prescribed by the
American National Standards Institute for electronic or automated
sphygmomanometers.
E-18