13
●
Function of counting scale F2.1 = 5
Instrument display
【
】
, showing current quantity.
Sampling method
1.
Check whether the scale is in zero, if not, press
『
Zero
』
for zero
setting.
2.
Place materials counted on the scale.
3. Press
『
Total
』
long until the display shows
【
】
, and then
press
『
』
. If F2.6 = 0, the display shows
【
】
. Input the quantity
counted just now, and press
『
』
for confirmation. The instrument saves
sampling data and quits the sampling interface. If F2.6 = 1, the display shows
【
】
, input sample weight. Press
『
』
, the instrument saves setting
data and quits sampling setting interface.
4. In this function, press
『
Hold
』
to display switch between quantity and
weight.
●
Function of animal scale F2.1 = 6
Operation method
In normal weighing state, place the animal on the weighing platform and
its weight must be ≥threshold value set in F2.2. Press
『
Total
』
, instrument
will collect data sampling. After sampling, the average value of sampling data
will be locked, showing A X.XXX. Press
『
』
to print; press
『
Hold
』
or
『
Total
』
to quit the interface.
●
VII Prompt Message of Instrument
The instrument has extremely high stability and reliability, thus is not easy to
具有极
have error in general situation. Once an error occurs, please make clear the error first
and observe whether the instrument still has error after power-on. Do not hurry to
repair the scale body or instrument. Repair the instrument according to error code of
the instrument as possible.
No.
Symbol
Analysis
Treatment Method
1
【
】
【
】
Unable to reset after startup
1.
Determine it is no-load state
in startup;
2.
Make zero calibration again.
Содержание LUW2001/LBW
Страница 1: ...User manual for LUW2001 LBW...