In order to rectify this error message, follow these steps;
a)
Place more sample on the weighing platter.
b)
Press the numerical (0-9) key to insert the new number of samples on the
weighing platter and press [WT/SET] to confirm the weight.
c)
Repeat step (a) and (b) until the ‘lack of sample’ sign disappear.
10.3
: Upper and lower limit alarm
This function is designed for packing purpose. For example, if the operator wishes to
perform a sample count of 1000 pieces for every package, he can activate the alarm
function of the scale when the sample count falls below or exceed the preset limits.
a)
Press [QTY/SET] button and to enter into the quantity alarm menu.
b)
Press [ZERO/MENU] button to toggle between the alarm ‘On’ or ‘Off’
function, where scale displays “CH=On” when the quantity alarm is activate
and “CH=OFF” when quantity alarm is deactivated.
c)
When the scale displays will show as per below diagram with a blinking ‘0’, set
the upper limit (H) of sample count to 1010 and the lower limit (L) of sample
count to 990.
d)
The scale displays will show as per below diagram, press [TARE/CHG] to confirm
the setting and exit the quantity alarm menu.