Page 14
Instruction Manual V1.3
7). Pulse Duty Fault
8). Product of V.I.T fault (Vd Fault)
9). Interlock.
After removing the fault conditions, press RESET button. If there is no fault
condition exists, the unit resumes the operation.
1.7.11 Vd Fault
This is a disabled function for HVC-XXXX/USB series models.
1.7.12 RESET
To reset the unit from the fault condition, this button is pressed. If there is
fault condition exists,
pressing RESET doesn’t recover the unit. First remove
the fault condition, and then press RESET.
1.7.13 ON/OFF/S.SHOT Output Switch
Output Switch OFF required before change:
1). CW/Pulse Selection
2). Output Device Selection
3). Output Voltage Level Selection
This button controls LD driver output both in CW and Pulse mode. Pressing
ON enables the LD driver to send the output current. Once ON button is
pressed, it stays on until OFF button is pressed.
In Pulse mode, the on time is not synchronized with the setting pulse. To
synchronize to the pulse, go to PULSE mode and use Run/Stop button on
pulse setting screen, and/or an external trigger. The detail is explained in a
separate pulse setting manual.
S.SHOT (Single Shot) enables the output current while a mouse left button is
pressed and disables when the mouse button is released. This works both in
CW and Pulse mode, but again it is not synchronized with the setting pulse in
Pulse mode. To synchronize a single shot to the setting pulse, go to pulse
setting screen, and use a single shot available in the screen.
1.8.1 Pulse Setting Screen
The detail of this screen and how to set the pulse combination are explained in a
separate manual. Refer to
attached with our products.
Содержание HVC-XXXX/QCW-USB