Section 3. Troubleshooting
Note:
Do not open the LI-870 enclosure in the field. This can cause the tem-
perature and pressure of the optics to become destabilized. More importantly, this
can allow dust and debris into your analyzer. If possible, you should confirm
whether your analyzer is working properly and troubleshoot any issues before tak-
ing it out to the field.
If your Smart Chamber or LI-8250 is on and your cables are properly connected,
the light should be blinking. If it is not blinking, something could be wrong with
your power cable, the chamber power-out connector, analyzer power-in connector,
or other electronics inside the case. Check for dust and debris in the connectors
and for damage to the cable itself. For power-related issues, it is recommended that
you call LI-COR technical support for further instructions.
If the light is blinking but you still cannot identify your device through the Smart
Chamber or LI-8250 software, check your USB cable for damage. Is there dust or
debris in the cable heads or the connectors? Try refreshing the interface software,
clearing your browser cache, and rebooting your Smart Chamber or LI-8250. If you
still cannot connect, contact LI-COR technical support.
Unable to Zero or Span the Instrument
Is there dirt in the optics? If the optical cell becomes contaminated, the instrument
will drift in either the zero or span. See
details on cleaning your optical bench.
Instrument Reports -50 ppm CO
2
or Measurements Jump Around
If the instrument measures -50 ppm or the measurements are going between neg-
ative and positive values, or just simply not making any sense, the optical source
may have failed or be in the midst of failure. Contact technical support for addi-
tional troubleshooting help.
3-2
Using the LI-870 CO₂/H₂O Analyzer
Содержание LI-870
Страница 1: ...Using the LI 870 CO2 H2O Analyzer...
Страница 2: ......
Страница 12: ...Section 1 Overview of the LI 870 1 4 Using the LI 870 CO H O Analyzer...
Страница 18: ...Section 2 Using the LI 870 2 6 Using the LI 870 CO H O Analyzer...
Страница 30: ...Appendix A Specifications A 2 Using the LI 870 CO H O Analyzer...
Страница 41: ......