Calibration Issues
GaAsP Light Sensors
Using the LI-6400
18-29
3
Orient the leaf chamber
Orient it so that the incoming radiation is perpendicular to the leaf plane. The
distance from the leaf chamber to the light source should not be changed be-
tween light sensor calibration and photosynthesis measurements. This will
minimize errors due to radiation geometry and view factors.
4
Set the Light Source to
Sun + Sky
This will apply no correction factor for your calibration readings.
5
Record the readings
Note the PAR values
Q
c
(ParIn_µm) and
Q
x
(ParOutµm)
6
Compute the correction factor
The correction factor
f
a
is
(18-1)
7
Enter the new value
Select
"Light Source Control"
in the Config Menu, press
labels
, then press
Edit StdFile
(
F2
). Add your new correction factor to the list, as shown in
Figure 18-23. YouÕll need a name in double quotes, followed by the
f
a
value,
followed by a guess at the absorbed energy conversion factor (use 0.2 unless
you know better).
Note: We do not recommend using an LI-190 quantum sensor to calibrate, or
even accurately measure, the LED light source output because the drop off in
quantum sensor sensitivity near 700nm (Figure 8-4 on page 8-8) is similar to
f
a
Q
x
Q
c
------
=
"Sun+Sky" 1.0 0.19
"Quartz Halogen" 1.11 0.26
"Fluorescent" 0.88 0.18
"Metal Halide" 0.87 0.17
"Neon Beer Sign" 0.70 .2_
/User/Configs/LightSources
Figure 18-23. Each entry in the list source list needs a quoted name, an actinity cor-
rection factor, and a conversion factor (used only for energy balance computations).
Light source name
(in quotes)
Correction factor (
f
a
)
Conversion factor for energy
balance computations
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...