background image

LAI-2050 Log Switch Assembly

Replacement Instructions for the LAI-2000 Plant Canopy Analyzer

1

4

Loosen the 4 hex
head screws; you
will need to pivot
the optical sensor
head to access all
of the screws.

The following document describes the procedure for
replacing the log switch assembly on the LAI-2000
optical sensor head (p/n 2000-906).  Please note that
the replacement kit does not include some required
adhesives; you can search for local suppliers and/or
order adhesives at the web addresses shown below.
Note, too, that the optical sensor head is a static-
sensitive device, and precautions (e.g. a static wrist
strap) should be taken before beginning this proce-
dure.

Tools Required:

3/32” hex wrench
1/4” open end wrench
Needle nose and lock-joint pliers
Flat blade screwdriver
Adhesives, including Loctite 242 Threadlocker and
Loctite PRISM 401 (www.loctite.com), and Dow
Corning 732 Multi-purpose sealant
(www.dowcorning.com)

Time Required:

 Approximately 30 minutes

Follow these steps to replace the log switch as-
sembly:

1.

Use the 3/32” hex wrench to loosen the 4 hex
head screws at the connection between the
handle and the optical sensor (Figure 1).  You will
need to pivot the sensor head to access the
screws.

2.

Use a pair of needle nose pliers to gently pull out
the detector connector from the circuit board
(Figure 2).

3.

Use the 3/32” hex wrench to remove the 4 hex
head screws at the opposite end of the handle,
where it attaches to the cable assembly (Figure 3).
Slide the circuit board out until you can unplug
the cable assembly by pulling the connector
straight out from the circuit board.  Continue to
slide the circuit board out until you can access the
4-pin (white) log button/grounding connector
(Figure 4).  Unplug the 4-pin connector, and slide
the circuit board back into the handle.

3

Remove the 4 hex
head screws at
the handle/cable
assembly.

2

Unplug the
detector connector
by pulling straight
out.

Slide the circuit
board out until you
can unplug the
white 4-pin
connector, and
then slide the
board back into the
handle.

Содержание LAI-2050

Страница 1: ...wcorning com Time Required Approximately 30 minutes Follow these steps to replace the log switch as sembly 1 Use the 3 32 hex wrench to loosen the 4 hex head screws at the connection between the handle and the optical sensor Figure 1 You will need to pivot the sensor head to access the screws 2 Use a pair of needle nose pliers to gently pull out the detector connector from the circuit board Figure...

Страница 2: ...the handle to remove the screw Remove the switch assembly Figure 7 Note The log switch has a small o ring that may adhere to the inside of the handle if it does simply use the screwdriver to pry it loose 6 Clean any RTV residue from the ends of the handle and the caps that mate to the handle tube Figure 8 Slide the circuit board out of the handle at the end opposite the log switch 6 Pry up the log...

Страница 3: ...d wrench as you tighten the screw Figure 2 4 Insert the circuit board into the handle from the end opposite the log switch Figure 3 the leading edge of the board has the 16 pin connector that attaches to the cable assembly and the board components face the same direction as the log switch Slide the board through the handle so that it extends through the opposite end far enough so that you can atta...

Страница 4: ...7 At tach the optical sensor to the handle using the 4 hex screws note the orientation of the handle and log switch in relation to the sensor head Figure 9 7 Apply a small drop of Loctite 401 to the top of the switch and install the button cap Figure 8 be careful not to apply too much adhesive as excess adhesive will lock the switch in place 6 Plug the sensor head connector into the circuit board ...

Отзывы: