6
ENGLISH
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz – 80 MHz
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
Portable and mobile RF communications equipment should be used no closer to any part of the EUT,
including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 2.7 GHz
Where P is the maximum output power rating of the transmitter in watts(W) according to the transmitter manufacturer and
d is the recommended separation distance in meters(M).
Field strengths from fixed RF transmitters as determined by an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Radiated RF EM fields
IEC 61000-4-3
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz