CALIBRATION SOFTWARE
When acquiring images from the detector, calibration is essential to obtain images of high quality.
enables you to create and check the necessary values for the calibration.
NOTE
•
It is recommended to perform a calibration once per month for the following three months after the purchase, and then once every six month to ensure the quality of images.
•
It is also recommended to turn on the detector for 15 minutes before the Calibration.
•
The default values set in
can be changed depending on the actual conditions of use.
Security
cannot be used independently without being connected to the detector. The software cannot perform all actions, including moving to another menu and confirming settings, without the actual
connection. In addition, even if the software is connected to the detector,
cannot be performed before the initial date of the product installation is registered.
Calibration Software
The features of the
include (Settings), [Calibration], [User BPM], [Validation] and [EI].
Содержание 17HK700G-W
Страница 6: ...6 ENGLISH PART NAME AND FUNCTION Detector FRONT BACK 100kg ...
Страница 13: ...13 ENGLISH Dimension Detector Unit mm Front Side 100kg 460 0 460 0 100kg 15 6 ...
Страница 14: ...14 ENGLISH Control Box Unit mm Front Side Power Ethernet Ready Exposure 125 0 255 109 8 ...
Страница 19: ...19 ENGLISH EXP_REQ EXP_ACK TimingChart ...
Страница 23: ...23 ENGLISH 2 Right clickLocalAreaConnection andclick Properties ...
Страница 26: ......
Страница 27: ...17HK700G W www lg com SOFTWARE MANUAL ...
Страница 37: ...11 ENGLISH Image Functions 5 3 2 4 1 ...
Страница 80: ...54 ENGLISH 2 Click the Windows Firewall link 3 On the left side of the pane click the Advanced Settings link ...