Connections
23
GA 05.145/3.02 - 02/2001
2.6.4 Interface description
The frequency converter has an optional serial interface
RS 485. It is operated with the standardized USS proto-
col. The frequency converter can be set to addresses 0
to 15. Addresses over 15 are not supported.
For more detailed information concerning the USS pro-
tocol please contact Leybold.
The RS 485 bus should be connected as shown in Fig.
22.
The frequency converter can be configured through the
parameters shown in the parameter list.
As to the RS 232 and Profibus interfaces please enqui-
re.
Input data / status
Output data
Operating mode
Start/
Pump
Normal
Error is
Motor
Relay
Relay
LED
LED
stop
rotating
frequency
present
drive
NORMAL
ERROR
STATUS
ERROR
signal
≥
90% of
OPERATION
setpoint
frequency
Stop
no
no
no
off
passive
passive
off
off
Pump not operating
Stop
yes
no
no
off
passive
passive
flashes
off
Pump is decelerating
Stop
yes
yes
no
off
passive
passive
flashes
off
Just after stop; pump was in the normal
operating mode before that
Start
no
no
no
on
passive
passive
off
off
Just after start
Start
yes
no
no
on
passive
passive
flashes
off
Pump is accelerating
Start
yes
yes
no
on
active
passive
green
off
Pump is in the normal operating mode
Stop
no
no
yes
off
passive
active
off
red
Error is present; pump is at standstill
Stop
yes
no
yes
off
passive
active
flashes
red
Error is present; pump is decelerating
Stop
yes
yes
yes
off
passive
active
flashes
red
Error has just occurred
Start
no
no
yes
off
passive
active
off
red
Error is present; pump is at standstill
Start
yes
no
yes
off
passive
active
flashes
red
Error is present; pump is decelerating
Start
yes
yes
yes
off
passive
active
flashes
red
Error has just occurred
Relay status
Other modes are not possible; they indicate
a failure affecting the TURBO.DRIVE S.
7
8
9
1
2
3
4
5
6
TxD/RxD +
TxD/RxD –
0,5 A, 24 V DC
Fig. 21 Pin assignment of the interface connector „RS 485“
Links for activation of
the bus terminator