background image

GA300553102_002_C2 

– 04/2017

 

 

7

 

 

3. 

Installation 

3.1 

Mechanical Installation 

 

The GRAPHIX CONTROLLER can be used as follows: 

 

  Rack installation  

  Front panel cut-out installation  

  Benchtop instrument 

 

 

WARNING: Powering down 

Install the instrument or place it so that you are in  a position to operate the 
mains  power  switch  at  any  time  or  to  ensure  that  the  instrument  can  be 
deenergised at any time. 
 

 

3.2 

Connections 

Figure 1, page 7 depicts the connection options on the rear of the GRAPHIX controller. 

 

 

Figure 1 

– 

Rear of the instrument 

 

Mains power connection with mains switch 
and instrument fuses 

Earth connection 

C1/C2  Connection of measurement channel 1 for 

sensors  (CH1 A und CH1 B) 

D1/D2  Connection of measurement channel 2 (for 

GRAPHIX TWO and THREE only) for 
sensors  (CH2 A and CH2 B)  

E1/E2  Connection of measurement channel 3 (for 

GRAPHIX THREE only) for sensors (CH3 A 
and CH3 B)  

Relay output connection (Relay Output) 

Analogue output, chart recorder output and 
external control connection (Control)  

RS232 or RS485 interface connection 
(RS232 / RS485) 

 

 

 

CAUTION: Impermissible Sensors. 

Connecting  sensors  which  have  not  been  designed  to  be  operated  in 
connection with the GRAPHIX controller or which do not comply with current 
EMC  guidelines  can  impair  operation  of  the  instrument  or  even  damage  it. 
Always  operate  the  GRAPHIX  controller  with  approved  sensors. 



 Chapter 1.2 

 Conforming Utilisation, page 4. 

 

 

CAUTION: Multiple Sensors. 

Only  a  single  sensor  may  be  connected  to  each  measurement  channel. 
Otherwise  the  connected  sensors  will  suffer  damage.  Connect  to  each 
measurement channel precisely one sensor only. 

 
 
 

Содержание 230680V01

Страница 1: ...Catalog No 230680V01 230681V01 230682V01 GRAPHIXONE GRAPHIXTWO GRAPHIXTHREE Vacuum Gauge Controller Operation Manual GA300553102_002_C2...

Страница 2: ......

Страница 3: ...Symbols 10 4 4 1 Symbols for the Controls 10 4 4 2 Symbols for the Language Selection 11 4 4 3 Status and Warning Symbols 11 4 5 Measuring Mode 12 5 Maintenance and Service 13 5 1 Maintenance 13 5 2 C...

Страница 4: ...HIX controller is a display and operating unit with a graphic user interface for sensors with an analogue or digital interface of the company Leybold GmbH or from other manufacturers Depending on the...

Страница 5: ...sume any responsibility or warranty in case the operator or third persons do not observe the information given in this document do not use the product as intended modify the product in any way convers...

Страница 6: ...h electricity NOTICE General information pointing to further information respectively reference sections 2 3 Basic Safety Information During all work like installation maintenance and repair activitie...

Страница 7: ...measurement channel 2 for GRAPHIX TWO and THREE only for sensors CH2 A and CH2 B E1 E2 Connection of measurement channel 3 for GRAPHIX THREE only for sensors CH3 A and CH3 B F Relay output connection...

Страница 8: ...sible You may select between the following display modes Normal Default display mode which provides all important information on the connected sensors Chart Graphic display of the pressure history of...

Страница 9: ...e functions In this way you may configure your GRAPHIX controller and also utilise further functions offered by the instrument Channel Menu Buttons Briefly touching the desired channel invokes the cha...

Страница 10: ...Enabling display mode Leak Test Speedo Channel 1 Enabling display mode Speedo for channel 1 Speedo Channel 2 Enabling display mode Speedo for channel 2 Speedo Channel 3 Enabling display mode Speedo f...

Страница 11: ...ge RU Russian Polish Select menu language PL Polish Turkish Select menu language TR Turkish Table 4 Symbols for language selection 4 4 3 Status and Warning Symbols Symbol Designation Explanation Statu...

Страница 12: ...in specified measuring range Measuring value and additional warning symbol in the display mode Normal and Speedo or Measuring value displayed by a yellow font in the display modes Chart Big and Leak...

Страница 13: ...n the USB stick in the folder RECOVERY o After the reset process has been completed the instrument will automatically restart with the default parameters o The GRAPHIX controller will now be ready for...

Страница 14: ...Setup o The question Are you sure is displayed to ensure that you really want to run the reset When not wanting to run the reset tap the button to cancel the reset process o The main menu is then disp...

Страница 15: ...yed again To run the update tap the button Start update o On the memory at the USB interface is installed the folder RECOVERY o Configuration data are stored in a file with the format rescue txt on th...

Страница 16: ...measurement channel Acknowledge this message by selecting and terminating the channel menu at the corresponding channel Analog sensor signal out of range too high Error H 1 4 The measurement signal f...

Страница 17: ...memory stick Error usb file closing none 3 File on USB stick cannot be stored Check the attached USB memory stick Error usb write none 4 USB stick not recordably Check the attached USB memory stick Er...

Страница 18: ...n The setup value is saved To exit tap the button o The parameter selection display is displayed again 6 3 Help in Case of Malfunctions If the malfunction persists even after having exchanged the sens...

Страница 19: ...ry room During storage the following ambient conditions need to be maintained Ambient temperature 20 60 C Humidity of the air As low as possible Preferably in a sealed plastic bag with desiccant 7 3 W...

Страница 20: ...rvice zh leybold com Spain Leybold Spain S A C Huelva 7 E 08940 Cornell de Llobregat Barcelona Sales T 34 93 666 43 11 F 34 93 666 43 70 sales ba leybold com Service T 34 93 666 46 11 F 34 93 685 43 7...

Отзывы: