44
GA130001441_002_C0
– 09/2016
8.3.4
RSS (Read Set Point Status)
Reading the status of the switching thresholds for a channel.
S:
RSS[a]<CR>
E:
b[,][TAB]c<CR>
Parameter
Description
a
Channel number
1 = Channel 1 (TM1)
2 = Channel 2 (TM2)
3 = Channel 3 (PM)
b
Status SP1
0 = low
1 = high
c
Status SP2
0 = low
1 = high
8.3.5
SHV (Set HV on/off)
Setting HV on/off for PENNINGVAC, if switch-on and switch-off type have been set to
manual.
S:
SHV[a, b]<CR>
E:
OK<CR>
Parameter
Description
a
Channel number
3 = Channel 3 (PM)
b
HV on/off
0 = off
1 = on
8.3.6
SKL (Set Key Lock on/off)
Switching the keylock on or off.
S:
SKL[a]<CR>
E:
OK<CR>
Parameter
Description
a
Key lock
0 = off
1 = on