GA10.218_1.01 - 11/2004
6.4.6 Alarm delay
•
>
>
In some applications (for instance during pump down in a „chamber test
system“) it might be necessary to block an alarm for some time after pres-
sing START.
This delay time of the alarm can be changed.
After pressing START the loudspeaker is activated as soon as the leak
rate drops below trigger level 1 or after the entered alarm delay time has
elapsed. This setting is only active for the audio alarm types SETPOINT
and TRIGGER ALARM (See Chapter
).
Softkey 2: Arrow down
Decrease the minimum loudness. The minimum value is 0.
Softkey 3: Arrow down
Decrease the actual volume. The minimum value is limited by the mini-
mum volume.
Softkey 4: Beep off / Beep on
Softkey 5: Help function
Softkey 6: Arrow up
Increase the minimum volume. The maximum value is 15.
Softkey 7: Arrow up
Increase the regular volume. The maximum value is 15.
Softkey 3: Arrow down
Decrease the delay time. The minimum value is 0 seconds.
Softkey 5: Help function
Softkey 7: Arrow up
Increase the delay time. The maximum value is 10 minutes up to infinity.